ELISA Kit for Corticotropin Like Intermediate Lobe Peptide (CLIP)

Specificity

This assay has high sensitivity and excellent specificity for detection of Corticotropin Like Intermediate Lobe Peptide (CLIP).
No significant cross-reactivity or interference between Corticotropin Like Intermediate Lobe Peptide (CLIP) and analogues was observed.

Recovery

Matrices listed below were spiked with certain level of recombinant Corticotropin Like Intermediate Lobe Peptide (CLIP) and the recovery rates were calculated by comparing the measured value to the expected amount of Corticotropin Like Intermediate Lobe Peptide (CLIP) in samples.

Matrix Recovery range (%) Average(%)
serum(n=5) 89-97 93
EDTA plasma(n=5) 92-101 97
heparin plasma(n=5) 93-104 99

Precision

Intra-assay Precision (Precision within an assay): 3 samples with low, middle and high level Corticotropin Like Intermediate Lobe Peptide (CLIP) were tested 20 times on one plate, respectively.
Inter-assay Precision (Precision between assays): 3 samples with low, middle and high level Corticotropin Like Intermediate Lobe Peptide (CLIP) were tested on 3 different plates, 8 replicates in each plate.
CV(%) = SD/meanX100
Intra-Assay: CV<10%
Inter-Assay: CV<12%

Linearity

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of Corticotropin Like Intermediate Lobe Peptide (CLIP) and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.

Sample 1:2 1:4 1:8 1:16
serum(n=5) 97-104% 91-99% 92-99% 78-94%
EDTA plasma(n=5) 96-105% 78-102% 81-91% 99-105%
heparin plasma(n=5) 92-105% 84-101% 93-104% 84-104%

Stability

The stability of kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage condition.
To minimize extra influence on the performance, operation procedures and lab conditions, especially room temperature, air humidity, incubator temperature should be strictly controlled. It is also strongly suggested that the whole assay is performed by the same operator from the beginning to the end.

Reagents and materials provided

Reagents Quantity Reagents Quantity
Pre-coated, ready to use 96-well strip plate 1 Plate sealer for 96 wells 4
Standard 2 Standard Diluent 1×20mL
Detection Reagent A 1×120µL Assay Diluent A 1×12mL
Detection Reagent B 1×120µL Assay Diluent B 1×12mL
TMB Substrate 1×9mL Stop Solution 1×6mL
Wash Buffer (30 × concentrate) 1×20mL Instruction manual 1

Assay procedure summary

1. Prepare all reagents, samples and standards;
2. Add 50µL standard or sample to each well.
    And then add 50µL prepared Detection Reagent A immediately.
    Shake and mix. Incubate 1 hour at 37°C;
3. Aspirate and wash 3 times;
4. Add 100µL prepared Detection Reagent B. Incubate 30 minutes at 37°C;
5. Aspirate and wash 5 times;
6. Add 90µL Substrate Solution. Incubate 10-20 minutes at 37°C;
7. Add 50µL Stop Solution. Read at 450 nm immediately.

GIVEAWAYS

INCREMENT SERVICES

Magazine Citations
Multiple Sclerosis and Related Disorders Fatigue in Multiple Sclerosis: Is it related to cytokines and hypothalamic-pituitary-adrenal axis? S2211034817300494
Multiple Sclerosis and Related Disorders Fatigue in Multiple Sclerosis: is it related to cytokines and hypothalamic-pituitary-adrenal axis? pubmed:28641771
Catalog No. Related products for research use of Homo sapiens (Human) Organism species Applications (RESEARCH USE ONLY!)
CED064Hu ELISA Kit for Corticotropin Like Intermediate Lobe Peptide (CLIP) Enzyme-linked immunosorbent assay for Antigen Detection.
LMD064Hu Multiplex Assay Kit for Corticotropin Like Intermediate Lobe Peptide (CLIP) ,etc. by FLIA (Flow Luminescence Immunoassay) FLIA Kit for Antigen Detection.