ELISA Kit for Creatinine (CTN)

Cr

Specificity

This assay has high sensitivity and excellent specificity for detection of Creatinine (CTN).
No significant cross-reactivity or interference between Creatinine (CTN) and analogues was observed.

Recovery

Matrices listed below were spiked with certain level of Creatinine (CTN) and the recovery rates were calculated by comparing the measured value to the expected amount of Creatinine (CTN) in samples.

Matrix Recovery range (%) Average(%)
serum(n=5) 91-98 95
EDTA plasma(n=5) 94-105 101
heparin plasma(n=5) 92-104 97

Precision

Intra-assay Precision (Precision within an assay): 3 samples with low, middle and high level Creatinine (CTN) were tested 20 times on one plate, respectively.
Inter-assay Precision (Precision between assays): 3 samples with low, middle and high level Creatinine (CTN) were tested on 3 different plates, 8 replicates in each plate.
CV(%) = SD/meanX100
Intra-Assay: CV<10%
Inter-Assay: CV<12%

Linearity

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of Creatinine (CTN) and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.

Sample 1:2 1:4 1:8 1:16
serum(n=5) 80-102% 94-101% 94-102% 88-101%
EDTA plasma(n=5) 90-102% 90-98% 97-105% 83-97%
heparin plasma(n=5) 88-101% 82-92% 80-89% 86-104%

Stability

The stability of kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage condition.
To minimize extra influence on the performance, operation procedures and lab conditions, especially room temperature, air humidity, incubator temperature should be strictly controlled. It is also strongly suggested that the whole assay is performed by the same operator from the beginning to the end.

Reagents and materials provided

Reagents Quantity Reagents Quantity
Pre-coated, ready to use 96-well strip plate 1 Plate sealer for 96 wells 4
Standard 2 Standard Diluent 1×20mL
Detection Reagent A 1 Assay Diluent A 1×12mL
Detection Reagent B 1×120µL Assay Diluent B 1×12mL
Reagent Diluent 1×300µL Stop Solution 1×6mL
TMB Substrate 1×9mL Instruction manual 1
Wash Buffer (30 × concentrate) 1×20mL

Assay procedure summary

1. Prepare all reagents, samples and standards;
2. Add 50µL standard or sample to each well.
    And then add 50µL prepared Detection Reagent A immediately.
    Shake and mix. Incubate 1 hour at 37°C;
3. Aspirate and wash 3 times;
4. Add 100µL prepared Detection Reagent B. Incubate 30 minutes at 37°C;
5. Aspirate and wash 5 times;
6. Add 90µL Substrate Solution. Incubate 10-20 minutes at 37°C;
7. Add 50µL Stop Solution. Read at 450 nm immediately.

GIVEAWAYS

INCREMENT SERVICES

Magazine Citations
Nature Feeding induces cholesterol biosynthesis via the mTORC1–USP20–HMGCR axis Pubmed: 33177714
Journal of Comparative Physiology B-Biochemical Systems and Environmental Physiology The physiological and molecular mechanisms to maintain water and salt homeostasis in response to high salt intake in Mongolian gerbils (Meriones unguiculatus) Pubmed: 32556536
Biomed Research International An Inhibitor of DRP1 (Mdivi-1) Alleviates LPS-Induced Septic AKI by Inhibiting NLRP3 Inflammasome Activation Pubmed: 32733934
Nutrition Gut Dysbiosis Contribute to High Fructose Induced Salt Sensitive Hypertension In Sprague-Dawley Rats
Intensive Care Medicine Experimental Preservation of renal endothelial integrity and reduction of renal edema by aprotinin does not preserve renal perfusion and function following experimental … 34169407
Catalog No. Related products for research use of Pan-species (General) Organism species Applications (RESEARCH USE ONLY!)
CPV806Ge21 OVA Conjugated Creatinine (CTN) Immunogen; SDS-PAGE; WB.
CEV806Ge ELISA Kit for Creatinine (CTN) Enzyme-linked immunosorbent assay for Antigen Detection.
LMV806Ge Multiplex Assay Kit for Creatinine (CTN) ,etc. by FLIA (Flow Luminescence Immunoassay) FLIA Kit for Antigen Detection.