Multiplex Assay Kit for Beta-2-Microglobulin (b2M) ,etc. by FLIA (Flow Luminescence Immunoassay)

BMG; B2MG

(Note: Up to 8-plex in one testing reaction)

Specificity

This assay has high sensitivity and excellent specificity for detection of Beta-2-Microglobulin (b2M) ,etc. by FLIA (Flow Luminescence Immunoassay).
No significant cross-reactivity or interference between Beta-2-Microglobulin (b2M) ,etc. by FLIA (Flow Luminescence Immunoassay) and analogues was observed.

Recovery

Matrices listed below were spiked with certain level of recombinant Beta-2-Microglobulin (b2M) ,etc. by FLIA (Flow Luminescence Immunoassay) and the recovery rates were calculated by comparing the measured value to the expected amount of Beta-2-Microglobulin (b2M) ,etc. by FLIA (Flow Luminescence Immunoassay) in samples.

Matrix Recovery range (%) Average(%)
serum(n=5) 95-103 98
EDTA plasma(n=5) 82-96 86
heparin plasma(n=5) 91-105 98

Precision

Intra-assay Precision (Precision within an assay): 3 samples with low, middle and high level Beta-2-Microglobulin (b2M) ,etc. by FLIA (Flow Luminescence Immunoassay) were tested 20 times on one plate, respectively.
Inter-assay Precision (Precision between assays): 3 samples with low, middle and high level Beta-2-Microglobulin (b2M) ,etc. by FLIA (Flow Luminescence Immunoassay) were tested on 3 different plates, 8 replicates in each plate.
CV(%) = SD/meanX100
Intra-Assay: CV<10%
Inter-Assay: CV<12%

Linearity

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of Beta-2-Microglobulin (b2M) ,etc. by FLIA (Flow Luminescence Immunoassay) and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.

Sample 1:2 1:4 1:8 1:16
serum(n=5) 79-92% 89-104% 79-104% 98-105%
EDTA plasma(n=5) 96-104% 95-105% 99-105% 81-98%
heparin plasma(n=5) 95-103% 79-90% 96-105% 80-97%

Stability

The stability of kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage condition.
To minimize extra influence on the performance, operation procedures and lab conditions, especially room temperature, air humidity, incubator temperature should be strictly controlled. It is also strongly suggested that the whole assay is performed by the same operator from the beginning to the end.

Reagents and materials provided

Reagents Quantity Reagents Quantity
96-well plate 1 Plate sealer for 96 wells 4
Pre-Mixed Standard 2 Standard Diluent 1×20mL
Pre-Mixed Magnetic beads (22#:b2M) 1 Analysis buffer 1×20mL
Pre-Mixed Detection Reagent A 1×120μL Assay Diluent A 1×12mL
Detection Reagent B (PE-SA) 1×120μL Assay Diluent B 1×12mL
Sheath Fluid 1×10mL Wash Buffer (30 × concentrate) 1×20mL
Instruction manual 1

Assay procedure summary

1. Preparation of standards, reagents and samples before the experiment;
2. Add 100μL standard or sample to each well,
    add 10μL magnetic beads, and incubate 90min at 37°C on shaker;
3. Remove liquid on magnetic frame, add 100μL prepared Detection Reagent A. Incubate 60min at 37°C on shaker;
4. Wash plate on magnetic frame for three times;
5. Add 100μL prepared Detection Reagent B, and incubate 30 min at 37°C on shaker;
6. Wash plate on magnetic frame for three times;
7. Add 100μL sheath solution, swirl for 2 minutes, read on the machine.

GIVEAWAYS

INCREMENT SERVICES

Magazine Citations
BioMed Research International The Renal Effects of Vanadate Exposure: Potential Biomarkers and Oxidative Stress as a Mechanism of Functional Renal Disorders—Preliminary Studies Hindawi: 740105
BMC Nephrology Absence of chloride intracellular channel 4 (CLIC4) predisposes to acute kidney injury but has minimal impact on recovery Biomedcentral: Source
Nature Communications Mef2C restrains microglial inflammatory response and is lost in brain ageing in an IFN-I-dependent manner. pubmed:28959042
PLoS One Urine proteome analysis by C18 plate–matrix-assisted laser desorption/ionization time-of-flight mass spectrometry allows noninvasive differential diagnosis … Pubmed:30024955
Acta Anaesthesiologica Scandinavica Early prognostic factors in septic shock cancer patients: a prospective study with a proteomic approach Pubmed:29315472
Toxicology Mechanisms and Methods Long-term exposure to crotonaldehyde causes heart and kidney dysfunction through induction of inflammatory and oxidative damage in male Wistar rats Pubmed: 30426860
Journal of Clinical Medicine Urinary Proteomics for the Early Diagnosis of Diabetic Nephropathy in Taiwanese Patients Pubmed: 30486327
Journal of Proteomics iTRAQ-based analysis for the identification of MARCH8 targets in human esophageal squamous cell carcinoma 33540066
Antioxidants Diminishment of Nrf2 Antioxidative Defense Aggravates Nephrotoxicity of Melamine and Oxalate Coexposure. Antioxidants 2021, 10, 1464 34573096
Int Arch Occup Environ Health Biomarkers of cadmium exposure and renal function in estuarine adult villagers 34773507
Catalog No. Related products for research use of Rhesus monkey (Simian) Organism species Applications (RESEARCH USE ONLY!)
RPA260Si01 Recombinant Beta-2-Microglobulin (b2M) Positive Control; Immunogen; SDS-PAGE; WB.
PAA260Si01 Polyclonal Antibody to Beta-2-Microglobulin (b2M) WB; IHC; ICC; IP.
SEA260Si ELISA Kit for Beta-2-Microglobulin (b2M) Enzyme-linked immunosorbent assay for Antigen Detection.
LMA260Si Multiplex Assay Kit for Beta-2-Microglobulin (b2M) ,etc. by FLIA (Flow Luminescence Immunoassay) FLIA Kit for Antigen Detection.