Multiplex Assay Kit for Motilin (MTL) ,etc. by FLIA (Flow Luminescence Immunoassay)

MLN; MAP; Promotilin; Motilin-associated peptide

(Note: Up to 8-plex in one testing reaction)

Specificity

This assay has high sensitivity and excellent specificity for detection of Motilin (MTL) ,etc. by FLIA (Flow Luminescence Immunoassay).
No significant cross-reactivity or interference between Motilin (MTL) ,etc. by FLIA (Flow Luminescence Immunoassay) and analogues was observed.

Recovery

Matrices listed below were spiked with certain level of recombinant Motilin (MTL) ,etc. by FLIA (Flow Luminescence Immunoassay) and the recovery rates were calculated by comparing the measured value to the expected amount of Motilin (MTL) ,etc. by FLIA (Flow Luminescence Immunoassay) in samples.

Matrix Recovery range (%) Average(%)
serum(n=5) 79-93 88
EDTA plasma(n=5) 98-105 101
heparin plasma(n=5) 93-101 97

Precision

Intra-assay Precision (Precision within an assay): 3 samples with low, middle and high level Motilin (MTL) ,etc. by FLIA (Flow Luminescence Immunoassay) were tested 20 times on one plate, respectively.
Inter-assay Precision (Precision between assays): 3 samples with low, middle and high level Motilin (MTL) ,etc. by FLIA (Flow Luminescence Immunoassay) were tested on 3 different plates, 8 replicates in each plate.
CV(%) = SD/meanX100
Intra-Assay: CV<10%
Inter-Assay: CV<12%

Linearity

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of Motilin (MTL) ,etc. by FLIA (Flow Luminescence Immunoassay) and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.

Sample 1:2 1:4 1:8 1:16
serum(n=5) 78-98% 86-101% 93-104% 89-101%
EDTA plasma(n=5) 86-103% 79-95% 84-95% 93-105%
heparin plasma(n=5) 90-98% 88-101% 80-95% 82-91%

Stability

The stability of kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage condition.
To minimize extra influence on the performance, operation procedures and lab conditions, especially room temperature, air humidity, incubator temperature should be strictly controlled. It is also strongly suggested that the whole assay is performed by the same operator from the beginning to the end.

Reagents and materials provided

Reagents Quantity Reagents Quantity
96-well plate 1 Plate sealer for 96 wells 4
Pre-Mixed Standard 2 Standard Diluent 1×20mL
Pre-Mixed Magnetic beads (22#:MTL) 1 Analysis buffer 1×20mL
Pre-Mixed Detection Reagent A 1×120μL Assay Diluent A 1×12mL
Detection Reagent B (PE-SA) 1×120μL Assay Diluent B 1×12mL
Sheath Fluid 1×10mL Wash Buffer (30 × concentrate) 1×20mL
Instruction manual 1

Assay procedure summary

1. Preparation of standards, reagents and samples before the experiment;
2. Add 50μL standard or sample to each well,
    add 10μL magnetic beads,and 50μL Detection Reagent A,incubate 60min at 37°C on shaker;
3. Wash plate on magnetic frame for three times;
4. Add 100μL prepared Detection Reagent B, and incubate 30 min at 37°C on shaker;
5. Wash plate on magnetic frame for three times;
6. Add 100μL sheath solution, swirl for 2 minutes, read on the machine.

GIVEAWAYS

INCREMENT SERVICES

Magazine Citations
世界华人消化杂志 清醒状态下热射病大鼠胃肠动力与胃泌素、胃动素的变化 Wjgnet:Source
Sci Rep. Using the "target constituent removal combined with bioactivity assay" strategy to investigate the optimum arecoline content in charred areca nut. pubmed:28054652
Applied Biological Chemistry Positive enhancement of Lactobacillus fermentum HY01 on intestinal movements of mice having constipation 10.1007/s13765-017-0327-3
Biomed Research International Colonic Transit Time and Gut Peptides in Adult Patients with Slow and Normal Colonic Transit Constipation pubmed:29082242
Evidence-Based Complementary and Alternative Medicine Home-Based Transcutaneous Neuromodulation Improved Constipation via Modulating Gastrointestinal Hormones and Bile Acids Pubmed:29853946
The American Journal of the Medical Sciences The Relations Among Serum Ghrelin, Motilin and Gastric Emptying and Autonomic Function in Autoimmune Gastritis Pubmed:29753372
NEUROGASTROENTEROLOGY AND MOTILITY Integrative effects of transcutaneous electrical acustimulation on abdominal pain, gastrointestinal motility, and inflammation in patients with early‐stage acute … 34536258
Evid Based Complement Alternat Med The Inhibitory Effects of Naringin in a Rat Model of Postoperative Intraperitoneal Adhesion Formation Pubmed:35069760
Catalog No. Related products for research use of Rattus norvegicus (Rat) Organism species Applications (RESEARCH USE ONLY!)
RPA575Ra01 Recombinant Motilin (MTL) Positive Control; Immunogen; SDS-PAGE; WB.
PAA575Ra01 Polyclonal Antibody to Motilin (MTL) WB; IHC; ICC; IP.
MAA575Ra21 Monoclonal Antibody to Motilin (MTL) WB; IHC; ICC; IP.
CEA575Ra ELISA Kit for Motilin (MTL) Enzyme-linked immunosorbent assay for Antigen Detection.
LMA575Ra Multiplex Assay Kit for Motilin (MTL) ,etc. by FLIA (Flow Luminescence Immunoassay) FLIA Kit for Antigen Detection.
KSA575Ra11 ELISA Kit DIY Materials for Motilin (MTL) Main materials for "Do It (ELISA Kit) Yourself".