Multiplex Assay Kit for Hemoglobin (HB) ,etc. by FLIA (Flow Luminescence Immunoassay)

Hgb; Haemoglobin; Heterotetramer(αβ)2

(Note: Up to 8-plex in one testing reaction)

Specificity

This assay has high sensitivity and excellent specificity for detection of Hemoglobin (HB) ,etc. by FLIA (Flow Luminescence Immunoassay).
No significant cross-reactivity or interference between Hemoglobin (HB) ,etc. by FLIA (Flow Luminescence Immunoassay) and analogues was observed.

Recovery

Matrices listed below were spiked with certain level of recombinant Hemoglobin (HB) ,etc. by FLIA (Flow Luminescence Immunoassay) and the recovery rates were calculated by comparing the measured value to the expected amount of Hemoglobin (HB) ,etc. by FLIA (Flow Luminescence Immunoassay) in samples.

Matrix Recovery range (%) Average(%)
serum(n=5) 90-104 95
EDTA plasma(n=5) 87-95 91
heparin plasma(n=5) 98-105 101

Precision

Intra-assay Precision (Precision within an assay): 3 samples with low, middle and high level Hemoglobin (HB) ,etc. by FLIA (Flow Luminescence Immunoassay) were tested 20 times on one plate, respectively.
Inter-assay Precision (Precision between assays): 3 samples with low, middle and high level Hemoglobin (HB) ,etc. by FLIA (Flow Luminescence Immunoassay) were tested on 3 different plates, 8 replicates in each plate.
CV(%) = SD/meanX100
Intra-Assay: CV<10%
Inter-Assay: CV<12%

Linearity

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of Hemoglobin (HB) ,etc. by FLIA (Flow Luminescence Immunoassay) and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.

Sample 1:2 1:4 1:8 1:16
serum(n=5) 83-102% 83-101% 92-105% 78-96%
EDTA plasma(n=5) 78-96% 78-90% 89-102% 86-103%
heparin plasma(n=5) 95-104% 78-96% 81-105% 92-101%

Stability

The stability of kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage condition.
To minimize extra influence on the performance, operation procedures and lab conditions, especially room temperature, air humidity, incubator temperature should be strictly controlled. It is also strongly suggested that the whole assay is performed by the same operator from the beginning to the end.

Reagents and materials provided

Reagents Quantity Reagents Quantity
96-well plate 1 Plate sealer for 96 wells 4
Pre-Mixed Standard 2 Standard Diluent 1×20mL
Pre-Mixed Magnetic beads (22#:HB) 1 Analysis buffer 1×20mL
Pre-Mixed Detection Reagent A 1×120μL Assay Diluent A 1×12mL
Detection Reagent B (PE-SA) 1×120μL Assay Diluent B 1×12mL
Sheath Fluid 1×10mL Wash Buffer (30 × concentrate) 1×20mL
Instruction manual 1

Assay procedure summary

1. Preparation of standards, reagents and samples before the experiment;
2. Add 100μL standard or sample to each well,
    add 10μL magnetic beads, and incubate 90min at 37°C on shaker;
3. Remove liquid on magnetic frame, add 100μL prepared Detection Reagent A. Incubate 60min at 37°C on shaker;
4. Wash plate on magnetic frame for three times;
5. Add 100μL prepared Detection Reagent B, and incubate 30 min at 37°C on shaker;
6. Wash plate on magnetic frame for three times;
7. Add 100μL sheath solution, swirl for 2 minutes, read on the machine.

GIVEAWAYS

INCREMENT SERVICES

Magazine Citations
Comparative Biochemistry and Physiology Part C: Toxicology &amp; Pharmacology Effects of ectoine on behavioural, physiological and biochemical parameters of Daphnia magna Pubmed:25460046
PLoS One. Comparative Proteomics Identifies Host Immune System Proteins Affected by Infection with Mycobacterium bovis Pubmed:27027307
Parasites & Vectors The intracellular bacterium Anaplasma phagocytophilum selectively manipulates the levels of vertebrate host proteins in the tick vector Ixodes scapularis. pubmed:27561965
Proteomics. Clinical Applications Elevated levels of protein AMBP in cerebrospinal fluid of women with preeclampsia compared to normotensive pregnant women. pubmed:27615121
Molecular nutrition & food research Chicory inulin ameliorates type 2 diabetes mellitus and suppresses JNK and MAPK pathways in vivo and in vitro pubmed:28105758
Frontiers in Immunology Oxidized Hemoglobin Is Antigenic and Immunogenic in Lupus pubmed:28694810
新規人工酸素運搬体HemoAct を用いた脳虚血再灌流障害治療開発に関する研究 :
Acta Neuropathologica Communications Mural cell-derived laminin-α5 plays a detrimental role in ischemic stroke Pubmed: 30777135
Angiogenesis Decylubiquinone suppresses breast cancer growth and metastasis by inhibiting angiogenesis via the ROS/p53/BAI1 signaling pathway Pubmed: 32020421
Food Bioscience Water-extractable zinc protoporphyrin IX in Parma ham predominantly exists as complexes with hemoglobin and myoglobin
Photochem Photobiol Effects of Human Serum Albumin on the Fluorescence Intensity and Tumor Imaging Properties of IR‐780 Dye 34687567
Catalog No. Related products for research use of Capra hircus; Caprine (Goat) Organism species Applications (RESEARCH USE ONLY!)
PAB409Cp01 Polyclonal Antibody to Hemoglobin (HB) WB; IHC; ICC; IP.
SEB409Cp ELISA Kit for Hemoglobin (HB) Enzyme-linked immunosorbent assay for Antigen Detection.
LMB409Cp Multiplex Assay Kit for Hemoglobin (HB) ,etc. by FLIA (Flow Luminescence Immunoassay) FLIA Kit for Antigen Detection.