Multiplex Assay Kit for Glucagon Like Peptide 2 (GLP2) ,etc. by FLIA (Flow Luminescence Immunoassay)

(Note: Up to 8-plex in one testing reaction)

Specificity

This assay has high sensitivity and excellent specificity for detection of Glucagon Like Peptide 2 (GLP2) ,etc. by FLIA (Flow Luminescence Immunoassay).
No significant cross-reactivity or interference between Glucagon Like Peptide 2 (GLP2) ,etc. by FLIA (Flow Luminescence Immunoassay) and analogues was observed.

Recovery

Matrices listed below were spiked with certain level of recombinant Glucagon Like Peptide 2 (GLP2) ,etc. by FLIA (Flow Luminescence Immunoassay) and the recovery rates were calculated by comparing the measured value to the expected amount of Glucagon Like Peptide 2 (GLP2) ,etc. by FLIA (Flow Luminescence Immunoassay) in samples.

Matrix Recovery range (%) Average(%)
serum(n=5) 80-98 91
EDTA plasma(n=5) 81-105 94
heparin plasma(n=5) 78-98 90

Precision

Intra-assay Precision (Precision within an assay): 3 samples with low, middle and high level Glucagon Like Peptide 2 (GLP2) ,etc. by FLIA (Flow Luminescence Immunoassay) were tested 20 times on one plate, respectively.
Inter-assay Precision (Precision between assays): 3 samples with low, middle and high level Glucagon Like Peptide 2 (GLP2) ,etc. by FLIA (Flow Luminescence Immunoassay) were tested on 3 different plates, 8 replicates in each plate.
CV(%) = SD/meanX100
Intra-Assay: CV<10%
Inter-Assay: CV<12%

Linearity

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of Glucagon Like Peptide 2 (GLP2) ,etc. by FLIA (Flow Luminescence Immunoassay) and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.

Sample 1:2 1:4 1:8 1:16
serum(n=5) 88-95% 82-98% 98-105% 85-98%
EDTA plasma(n=5) 88-95% 96-105% 95-103% 96-104%
heparin plasma(n=5) 96-105% 96-103% 94-105% 92-99%

Stability

The stability of kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage condition.
To minimize extra influence on the performance, operation procedures and lab conditions, especially room temperature, air humidity, incubator temperature should be strictly controlled. It is also strongly suggested that the whole assay is performed by the same operator from the beginning to the end.

Reagents and materials provided

Reagents Quantity Reagents Quantity
96-well plate 1 Plate sealer for 96 wells 4
Pre-Mixed Standard 2 Standard Diluent 1×20mL
Pre-Mixed Magnetic beads (22#:GLP2) 1 Analysis buffer 1×20mL
Pre-Mixed Detection Reagent A 1×120μL Assay Diluent A 1×12mL
Detection Reagent B (PE-SA) 1×120μL Assay Diluent B 1×12mL
Sheath Fluid 1×10mL Wash Buffer (30 × concentrate) 1×20mL
Instruction manual 1

Assay procedure summary

1. Preparation of standards, reagents and samples before the experiment;
2. Add 50μL standard or sample to each well,
    add 10μL magnetic beads,and 50μL Detection Reagent A,incubate 60min at 37°C on shaker;
3. Wash plate on magnetic frame for three times;
4. Add 100μL prepared Detection Reagent B, and incubate 30 min at 37°C on shaker;
5. Wash plate on magnetic frame for three times;
6. Add 100μL sheath solution, swirl for 2 minutes, read on the machine.

GIVEAWAYS

INCREMENT SERVICES

Magazine Citations
European Journal of Nutrition Improvement in glucose tolerance and insulin sensitivity by probiotic strains of Indian gut originin high-fat diet-fed C57BL/6J mice. pubmed:27757592
Scientific Reports Promising Biomarkers of Environmental Enteric Dysfunction: A Prospective Cohort study in Pakistani Children Pubmed:29445110
Journal of Lipid Research Novel GPR120 agonist TUG891 modulates fat taste perception and preference and activates tongue-brain-gut axis in mice Pubmed: 31806728
Gastroenterology Mucosal Genomics Implicate Lymphocyte Activation and Lipid Metabolism in Refractory Environmental Enteric Dysfunction 33524399
Catalog No. Related products for research use of Sus scrofa; Porcine (Pig) Organism species Applications (RESEARCH USE ONLY!)
CED059Po ELISA Kit for Glucagon Like Peptide 2 (GLP2) Enzyme-linked immunosorbent assay for Antigen Detection.
LMD059Po Multiplex Assay Kit for Glucagon Like Peptide 2 (GLP2) ,etc. by FLIA (Flow Luminescence Immunoassay) FLIA Kit for Antigen Detection.