Multiplex Assay Kit for Interleukin 18 Binding Protein (IL18BP) ,etc. by FLIA (Flow Luminescence Immunoassay)

IL18BPa; MC51L-53L-54L Homolog Gene Product; Tadekinig-alfa

(Note: Up to 8-plex in one testing reaction)

Specificity

This assay has high sensitivity and excellent specificity for detection of Interleukin 18 Binding Protein (IL18BP) ,etc. by FLIA (Flow Luminescence Immunoassay).
No significant cross-reactivity or interference between Interleukin 18 Binding Protein (IL18BP) ,etc. by FLIA (Flow Luminescence Immunoassay) and analogues was observed.

Recovery

Matrices listed below were spiked with certain level of recombinant Interleukin 18 Binding Protein (IL18BP) ,etc. by FLIA (Flow Luminescence Immunoassay) and the recovery rates were calculated by comparing the measured value to the expected amount of Interleukin 18 Binding Protein (IL18BP) ,etc. by FLIA (Flow Luminescence Immunoassay) in samples.

Matrix Recovery range (%) Average(%)
serum(n=5) 78-98 92
EDTA plasma(n=5) 88-104 101
heparin plasma(n=5) 79-90 83

Precision

Intra-assay Precision (Precision within an assay): 3 samples with low, middle and high level Interleukin 18 Binding Protein (IL18BP) ,etc. by FLIA (Flow Luminescence Immunoassay) were tested 20 times on one plate, respectively.
Inter-assay Precision (Precision between assays): 3 samples with low, middle and high level Interleukin 18 Binding Protein (IL18BP) ,etc. by FLIA (Flow Luminescence Immunoassay) were tested on 3 different plates, 8 replicates in each plate.
CV(%) = SD/meanX100
Intra-Assay: CV<10%
Inter-Assay: CV<12%

Linearity

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of Interleukin 18 Binding Protein (IL18BP) ,etc. by FLIA (Flow Luminescence Immunoassay) and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.

Sample 1:2 1:4 1:8 1:16
serum(n=5) 80-91% 89-98% 90-104% 88-95%
EDTA plasma(n=5) 80-91% 79-92% 90-97% 82-101%
heparin plasma(n=5) 92-105% 93-101% 80-88% 83-95%

Stability

The stability of kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage condition.
To minimize extra influence on the performance, operation procedures and lab conditions, especially room temperature, air humidity, incubator temperature should be strictly controlled. It is also strongly suggested that the whole assay is performed by the same operator from the beginning to the end.

Reagents and materials provided

Reagents Quantity Reagents Quantity
96-well plate 1 Plate sealer for 96 wells 4
Pre-Mixed Standard 2 Standard Diluent 1×20mL
Pre-Mixed Magnetic beads (22#:IL18BP) 1 Analysis buffer 1×20mL
Pre-Mixed Detection Reagent A 1×120μL Assay Diluent A 1×12mL
Detection Reagent B (PE-SA) 1×120μL Assay Diluent B 1×12mL
Sheath Fluid 1×10mL Wash Buffer (30 × concentrate) 1×20mL
Instruction manual 1

Assay procedure summary

1. Preparation of standards, reagents and samples before the experiment;
2. Add 100μL standard or sample to each well,
    add 10μL magnetic beads, and incubate 90min at 37°C on shaker;
3. Remove liquid on magnetic frame, add 100μL prepared Detection Reagent A. Incubate 60min at 37°C on shaker;
4. Wash plate on magnetic frame for three times;
5. Add 100μL prepared Detection Reagent B, and incubate 30 min at 37°C on shaker;
6. Wash plate on magnetic frame for three times;
7. Add 100μL sheath solution, swirl for 2 minutes, read on the machine.

GIVEAWAYS

INCREMENT SERVICES

Magazine Citations
Acta Physiologica IL-18 neutralization during alveolar hypoxia improves left ventricular diastolic function in mice Pubmed:25182570
Journal of Molecular Cell Biology Role of IL‐18 in atopic asthma is determined by balance of IL‐18/IL‐18BP/IL‐18R Pubmed:28922563
Scandinavian Journal of Immunology  Altered expression of IL‐18 binding protein and IL‐18 receptor in basophils and mast cells of asthma patients Pubmed:29505668
Scientific Reports IL-18 binding protein (IL-18BP) as a novel radiation countermeasure after radiation exposure in mice Pubmed: 33129789
Catalog No. Related products for research use of Homo sapiens (Human) Organism species Applications (RESEARCH USE ONLY!)
RPH821Hu01 Recombinant Interleukin 18 Binding Protein (IL18BP) Positive Control; Immunogen; SDS-PAGE; WB.
PAH821Hu01 Polyclonal Antibody to Interleukin 18 Binding Protein (IL18BP) WB; IHC; ICC; IP.
MAH821Hu21 Monoclonal Antibody to Interleukin 18 Binding Protein (IL18BP) WB; IHC; ICC; IP.
SEH821Hu ELISA Kit for Interleukin 18 Binding Protein (IL18BP) Enzyme-linked immunosorbent assay for Antigen Detection.
SCH821Hu CLIA Kit for Interleukin 18 Binding Protein (IL18BP) Chemiluminescent immunoassay for Antigen Detection.
LMH821Hu Multiplex Assay Kit for Interleukin 18 Binding Protein (IL18BP) ,etc. by FLIA (Flow Luminescence Immunoassay) FLIA Kit for Antigen Detection.