Multiplex Assay Kit for Cystatin B (CSTB) ,etc. by FLIA (Flow Luminescence Immunoassay)

Cys-B; CST6; EPM1; PME; STFB; CPI-B; Stefin B; Liver thiol proteinase inhibitor

(Note: Up to 8-plex in one testing reaction)

Specificity

This assay has high sensitivity and excellent specificity for detection of Cystatin B (CSTB) ,etc. by FLIA (Flow Luminescence Immunoassay).
No significant cross-reactivity or interference between Cystatin B (CSTB) ,etc. by FLIA (Flow Luminescence Immunoassay) and analogues was observed.

Recovery

Matrices listed below were spiked with certain level of recombinant Cystatin B (CSTB) ,etc. by FLIA (Flow Luminescence Immunoassay) and the recovery rates were calculated by comparing the measured value to the expected amount of Cystatin B (CSTB) ,etc. by FLIA (Flow Luminescence Immunoassay) in samples.

Matrix Recovery range (%) Average(%)
serum(n=5) 78-92 81
EDTA plasma(n=5) 80-95 80
heparin plasma(n=5) 95-105 101

Precision

Intra-assay Precision (Precision within an assay): 3 samples with low, middle and high level Cystatin B (CSTB) ,etc. by FLIA (Flow Luminescence Immunoassay) were tested 20 times on one plate, respectively.
Inter-assay Precision (Precision between assays): 3 samples with low, middle and high level Cystatin B (CSTB) ,etc. by FLIA (Flow Luminescence Immunoassay) were tested on 3 different plates, 8 replicates in each plate.
CV(%) = SD/meanX100
Intra-Assay: CV<10%
Inter-Assay: CV<12%

Linearity

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of Cystatin B (CSTB) ,etc. by FLIA (Flow Luminescence Immunoassay) and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.

Sample 1:2 1:4 1:8 1:16
serum(n=5) 80-92% 96-104% 87-96% 80-89%
EDTA plasma(n=5) 96-105% 79-92% 85-98% 91-101%
heparin plasma(n=5) 85-99% 89-101% 79-94% 83-96%

Stability

The stability of kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage condition.
To minimize extra influence on the performance, operation procedures and lab conditions, especially room temperature, air humidity, incubator temperature should be strictly controlled. It is also strongly suggested that the whole assay is performed by the same operator from the beginning to the end.

Reagents and materials provided

Reagents Quantity Reagents Quantity
96-well plate 1 Plate sealer for 96 wells 4
Pre-Mixed Standard 2 Standard Diluent 1×20mL
Pre-Mixed Magnetic beads (22#:CSTB) 1 Analysis buffer 1×20mL
Pre-Mixed Detection Reagent A 1×120μL Assay Diluent A 1×12mL
Detection Reagent B (PE-SA) 1×120μL Assay Diluent B 1×12mL
Sheath Fluid 1×10mL Wash Buffer (30 × concentrate) 1×20mL
Instruction manual 1

Assay procedure summary

1. Preparation of standards, reagents and samples before the experiment;
2. Add 100μL standard or sample to each well,
    add 10μL magnetic beads, and incubate 90min at 37°C on shaker;
3. Remove liquid on magnetic frame, add 100μL prepared Detection Reagent A. Incubate 60min at 37°C on shaker;
4. Wash plate on magnetic frame for three times;
5. Add 100μL prepared Detection Reagent B, and incubate 30 min at 37°C on shaker;
6. Wash plate on magnetic frame for three times;
7. Add 100μL sheath solution, swirl for 2 minutes, read on the machine.

GIVEAWAYS

INCREMENT SERVICES

Magazine Citations
AIDS. Cathepsin B and cystatin B in HIV-seropositive women are associated with infection and HIV-1-associated neurocognitive disorders PubMed: 23291538
International Journal of Circumpolar Health Evaluation of serum procathepsin B, cystatin B and cystatin C as possible biomarkers of ovarian cancer PubMed: PMC3754495
PLoS One. Dysregulation of Macrophage-Secreted Cathepsin B Contributes to HIV-1-Linked Neuronal Apoptosis PubMed: PMC3365072
Journal of Neuroimmune Pharmacology Cocaine potentiates cathepsin B secretion and neuronal apoptosis from HIV-infected macrophages Pubmed:25209871
European Journal of Cancer Supplements A91: Endogenous inhibitors of cysteine proteases and preform of cathepsin B in cancer of reproductive system Science: Article
Annals of Clinical Biochemistry Diagnostic specificity and sensitivity of PIVKAII, GP3, CSTB, SCCA1 and HGF for the diagnosis of hepatocellular carcinoma in patients with alcoholic liver cirrhosis. pubmed:28766361
Catalog No. Related products for research use of Homo sapiens (Human) Organism species Applications (RESEARCH USE ONLY!)
RPJ322Hu01 Recombinant Cystatin B (CSTB) Positive Control; Immunogen; SDS-PAGE; WB.
PAJ322Hu01 Polyclonal Antibody to Cystatin B (CSTB) WB;IHC;ICC/IF;FCM
MAJ322Hu21 Monoclonal Antibody to Cystatin B (CSTB) WB; IHC; ICC; IP.
SEJ322Hu ELISA Kit for Cystatin B (CSTB) Enzyme-linked immunosorbent assay for Antigen Detection.
LMJ322Hu Multiplex Assay Kit for Cystatin B (CSTB) ,etc. by FLIA (Flow Luminescence Immunoassay) FLIA Kit for Antigen Detection.