Mini Samples ELISA Kit for Interleukin 15 (IL15)

Specificity

This assay has high sensitivity and excellent specificity for detection of Mini Samples Interleukin 15 (IL15).
No significant cross-reactivity or interference between Mini Samples Interleukin 15 (IL15) and analogues was observed.

Recovery

Matrices listed below were spiked with certain level of recombinant Mini Samples Interleukin 15 (IL15) and the recovery rates were calculated by comparing the measured value to the expected amount of Mini Samples Interleukin 15 (IL15) in samples.

Matrix Recovery range (%) Average(%)
serum(n=5) 94-101 97
EDTA plasma(n=5) 87-104 94
heparin plasma(n=5) 82-101 90

Precision

Intra-assay Precision (Precision within an assay): 3 samples with low, middle and high level Mini Samples Interleukin 15 (IL15) were tested 20 times on one plate, respectively.
Inter-assay Precision (Precision between assays): 3 samples with low, middle and high level Mini Samples Interleukin 15 (IL15) were tested on 3 different plates, 8 replicates in each plate.
CV(%) = SD/meanX100
Intra-Assay: CV<10%
Inter-Assay: CV<12%

Linearity

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of Mini Samples Interleukin 15 (IL15) and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.

Sample 1:2 1:4 1:8 1:16
serum(n=5) 80-104% 94-104% 78-101% 82-93%
EDTA plasma(n=5) 88-96% 91-98% 84-96% 95-102%
heparin plasma(n=5) 98-105% 88-96% 93-102% 84-96%

Stability

The stability of kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage condition.
To minimize extra influence on the performance, operation procedures and lab conditions, especially room temperature, air humidity, incubator temperature should be strictly controlled. It is also strongly suggested that the whole assay is performed by the same operator from the beginning to the end.

Reagents and materials provided

Reagents Quantity Reagents Quantity
Pre-coated, ready to use 96-well strip plate 1 Plate sealer for 96 wells 4
Standard 2 Standard Diluent 1×20mL
Detection Reagent A 1×60µL Assay Diluent A 1×6mL
Detection Reagent B 1×60µL Assay Diluent B 1×6mL
TMB Substrate 1×4.5mL Stop Solution 1×3mL
Wash Buffer (30 × concentrate) 1×10mL Instruction manual 1

Assay procedure summary

1. Prepare all reagents, samples and standards;
2. Add 25µL standard or sample to each well. Incubate 1 hour at 37°C;
3. Aspirate and add 25µL prepared Detection Reagent A. Incubate 1 hour at 37°C;
4. Aspirate and wash 3 times;
5. Add 25µL prepared Detection Reagent B. Incubate 30 minutes at 37°C;
6. Aspirate and wash 5 times;
7. Add 25µL Substrate Solution. Incubate 10-20 minutes at 37°C;
8. Add 20µL Stop Solution. Read at 450nm immediately.

GIVEAWAYS

INCREMENT SERVICES

Magazine Citations
Endocrine Expression of interleukin-15 and inflammatory cytokines in skeletal muscles of STZ-induced diabetic rats: effect of resistance exercise training Pubmed: 24006180
Endocrine Time course of IL-15 expression after acute resistance exercise in trained rats: effect of diabetes and skeletal muscle phenotype Pubmed:25522723
Med Sci Moni miR-15b is Downregulated in Myasthenia Gravis Patients and Directly Regulates the Expression of Interleukin-15 (IL-15) in Experimental Myasthenia Gravis Mice PubMed: 26087886
Exercise physiology Effects of intrinsic aerobic capacity, aging and physical activity on interleukin-15 protein level in serum and skeletal muscle Dspace: Handle
Biotechnology Letters Novel adjuvant for immunization against tuberculosis: DNA vaccine expressing Mycobacterium tuberculosis antigen 85A and interleukin-15 fusion product elicits strong immune responses in mice s10529-017-2342-1
Journal of Controlled Release Interleukin-15 and cisplatin co-encapsulated thermosensitive polypeptide hydrogels for combined immuno-chemotherapy. pubmed:28408199
Catalog No. Related products for research use of Mus musculus (Mouse) Organism species Applications (RESEARCH USE ONLY!)
RPA061Mu02 Recombinant Interleukin 15 (IL15) Positive Control; Immunogen; SDS-PAGE; WB.
RPA061Mu01 Recombinant Interleukin 15 (IL15) Positive Control; Immunogen; SDS-PAGE; WB.
PAA061Mu01 Polyclonal Antibody to Interleukin 15 (IL15) WB; IHC; ICC; IP.
PAA061Mu02 Polyclonal Antibody to Interleukin 15 (IL15) WB; IHC; ICC; IP.
LAA061Mu71 Biotin-Linked Polyclonal Antibody to Interleukin 15 (IL15) WB; IHC; ICC.
LAA061Mu81 FITC-Linked Polyclonal Antibody to Interleukin 15 (IL15) WB; IHC; ICC; IF.
SEA061Mu ELISA Kit for Interleukin 15 (IL15) Enzyme-linked immunosorbent assay for Antigen Detection.
MEA061Mu Mini Samples ELISA Kit for Interleukin 15 (IL15) Enzyme-linked immunosorbent assay for Antigen Detection.
LMA061Mu Multiplex Assay Kit for Interleukin 15 (IL15) ,etc. by FLIA (Flow Luminescence Immunoassay) FLIA Kit for Antigen Detection.
KSA061Mu01 ELISA Kit DIY Materials for Interleukin 15 (IL15) Main materials for "Do It(ELISA Kit) Yourself"