Mini Samples ELISA Kit for Interleukin 7 (IL7)

Pre-B-cell Growth Factor; Lymphopoietin-1

Specificity

This assay has high sensitivity and excellent specificity for detection of Mini Samples Interleukin 7 (IL7).
No significant cross-reactivity or interference between Mini Samples Interleukin 7 (IL7) and analogues was observed.

Recovery

Matrices listed below were spiked with certain level of recombinant Mini Samples Interleukin 7 (IL7) and the recovery rates were calculated by comparing the measured value to the expected amount of Mini Samples Interleukin 7 (IL7) in samples.

Matrix Recovery range (%) Average(%)
serum(n=5) 86-98 93
EDTA plasma(n=5) 96-105 102
heparin plasma(n=5) 97-105 101

Precision

Intra-assay Precision (Precision within an assay): 3 samples with low, middle and high level Mini Samples Interleukin 7 (IL7) were tested 20 times on one plate, respectively.
Inter-assay Precision (Precision between assays): 3 samples with low, middle and high level Mini Samples Interleukin 7 (IL7) were tested on 3 different plates, 8 replicates in each plate.
CV(%) = SD/meanX100
Intra-Assay: CV<10%
Inter-Assay: CV<12%

Linearity

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of Mini Samples Interleukin 7 (IL7) and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.

Sample 1:2 1:4 1:8 1:16
serum(n=5) 98-105% 84-91% 78-98% 86-99%
EDTA plasma(n=5) 78-105% 79-96% 81-99% 86-101%
heparin plasma(n=5) 85-99% 88-102% 96-103% 99-105%

Stability

The stability of kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage condition.
To minimize extra influence on the performance, operation procedures and lab conditions, especially room temperature, air humidity, incubator temperature should be strictly controlled. It is also strongly suggested that the whole assay is performed by the same operator from the beginning to the end.

Reagents and materials provided

Reagents Quantity Reagents Quantity
Pre-coated, ready to use 96-well strip plate 1 Plate sealer for 96 wells 4
Standard 2 Standard Diluent 1×20mL
Detection Reagent A 1×60µL Assay Diluent A 1×6mL
Detection Reagent B 1×60µL Assay Diluent B 1×6mL
TMB Substrate 1×4.5mL Stop Solution 1×3mL
Wash Buffer (30 × concentrate) 1×10mL Instruction manual 1

Assay procedure summary

1. Prepare all reagents, samples and standards;
2. Add 25µL standard or sample to each well. Incubate 1 hour at 37°C;
3. Aspirate and add 25µL prepared Detection Reagent A. Incubate 1 hour at 37°C;
4. Aspirate and wash 3 times;
5. Add 25µL prepared Detection Reagent B. Incubate 30 minutes at 37°C;
6. Aspirate and wash 5 times;
7. Add 25µL Substrate Solution. Incubate 10-20 minutes at 37°C;
8. Add 20µL Stop Solution. Read at 450nm immediately.

GIVEAWAYS

INCREMENT SERVICES

Magazine Citations
Experimental Gerontology Specific alterations in the circulating levels of the SIRT1, TLR4, and IL7 proteins in patients with dementia Pubmed:30071285
Cardiovascular Therapeutics Role of IL‐7 in Fusion of Rat Bone Marrow Mesenchymal Stem Cells with Cardiomyocytes in vitro and Improvement of Cardiac Function in vivo Pubmed: 30451388
Stem Cells and Development The reparative effects of human adipose-derived mesenchymal stem cells in the chemotherapy damaged thymus Pubmed: 30511904
Catalog No. Related products for research use of Rattus norvegicus (Rat) Organism species Applications (RESEARCH USE ONLY!)
RPA662Ra01 Recombinant Interleukin 7 (IL7) Positive Control; Immunogen; SDS-PAGE; WB.
RPA662Ra02 Recombinant Interleukin 7 (IL7) Positive Control; Immunogen; SDS-PAGE; WB.
PAA662Ra01 Polyclonal Antibody to Interleukin 7 (IL7) WB; IHC; ICC; IP.
MAA662Ra21 Monoclonal Antibody to Interleukin 7 (IL7) WB; IHC; ICC; IP.
MEA662Ra Mini Samples ELISA Kit for Interleukin 7 (IL7) Enzyme-linked immunosorbent assay for Antigen Detection.
SEA662Ra ELISA Kit for Interleukin 7 (IL7) Enzyme-linked immunosorbent assay for Antigen Detection.
LMA662Ra Multiplex Assay Kit for Interleukin 7 (IL7) ,etc. by FLIA (Flow Luminescence Immunoassay) FLIA Kit for Antigen Detection.