Mini Samples ELISA Kit for Procollagen III (PCIII)

PCOL-III; PC-III; PC3; Type III Procollagen

Specificity

This assay has high sensitivity and excellent specificity for detection of Mini Samples Procollagen III (PCIII).
No significant cross-reactivity or interference between Mini Samples Procollagen III (PCIII) and analogues was observed.

Recovery

Matrices listed below were spiked with certain level of recombinant Mini Samples Procollagen III (PCIII) and the recovery rates were calculated by comparing the measured value to the expected amount of Mini Samples Procollagen III (PCIII) in samples.

Matrix Recovery range (%) Average(%)
serum(n=5) 89-103 93
EDTA plasma(n=5) 78-93 82
heparin plasma(n=5) 90-99 95

Precision

Intra-assay Precision (Precision within an assay): 3 samples with low, middle and high level Mini Samples Procollagen III (PCIII) were tested 20 times on one plate, respectively.
Inter-assay Precision (Precision between assays): 3 samples with low, middle and high level Mini Samples Procollagen III (PCIII) were tested on 3 different plates, 8 replicates in each plate.
CV(%) = SD/meanX100
Intra-Assay: CV<10%
Inter-Assay: CV<12%

Linearity

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of Mini Samples Procollagen III (PCIII) and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.

Sample 1:2 1:4 1:8 1:16
serum(n=5) 84-92% 83-101% 96-105% 96-104%
EDTA plasma(n=5) 92-101% 80-92% 89-102% 81-101%
heparin plasma(n=5) 93-102% 84-94% 79-104% 83-97%

Stability

The stability of kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage condition.
To minimize extra influence on the performance, operation procedures and lab conditions, especially room temperature, air humidity, incubator temperature should be strictly controlled. It is also strongly suggested that the whole assay is performed by the same operator from the beginning to the end.

Reagents and materials provided

Reagents Quantity Reagents Quantity
Pre-coated, ready to use 96-well strip plate 1 Plate sealer for 96 wells 4
Standard 2 Standard Diluent 1×20mL
Detection Reagent A 1×60µL Assay Diluent A 1×6mL
Detection Reagent B 1×60µL Assay Diluent B 1×6mL
TMB Substrate 1×4.5mL Stop Solution 1×3mL
Wash Buffer (30 × concentrate) 1×10mL Instruction manual 1

Assay procedure summary

1. Prepare all reagents, samples and standards;
2. Add 25µL standard or sample to each well. Incubate 1 hour at 37°C;
3. Aspirate and add 25µL prepared Detection Reagent A. Incubate 1 hour at 37°C;
4. Aspirate and wash 3 times;
5. Add 25µL prepared Detection Reagent B. Incubate 30 minutes at 37°C;
6. Aspirate and wash 5 times;
7. Add 25µL Substrate Solution. Incubate 10-20 minutes at 37°C;
8. Add 20µL Stop Solution. Read at 450nm immediately.

GIVEAWAYS

INCREMENT SERVICES

Magazine Citations
Free Radical Biology & Medicine Silybin combined with phosphatidylcholine and vitamin E in patients with nonalcoholic fatty liver disease: A randomized controlled trial PubMed: 22343419
Intensive Care Medicine Type III procollagen as a biomarker of susceptibility to ARDS? Pubmed:25617263
European Journal of Gastroenterology & Hepatology Measurements of serum procollagen-III peptide and M30 do not improve the diagnostic accuracy of transient elastography for the detection of hepatic fibrosis in patients with nonalcoholic fatty liver disease PubMed: 25923941
Biomedicine & Pharmacotherapy Bicyclol ameliorates advanced liver diseases in murine models via inhibiting the IL-6/STAT3 signaling pathway Pubmed:35658240
Catalog No. Related products for research use of Mus musculus (Mouse) Organism species Applications (RESEARCH USE ONLY!)
MED087Mu Mini Samples ELISA Kit for Procollagen III (PCIII) Enzyme-linked immunosorbent assay for Antigen Detection.
SED087Mu ELISA Kit for Procollagen III (PCIII) Enzyme-linked immunosorbent assay for Antigen Detection.
LMD087Mu Multiplex Assay Kit for Procollagen III (PCIII) ,etc. by FLIA (Flow Luminescence Immunoassay) FLIA Kit for Antigen Detection.