CLIA Kit for Beta-Lactoglobulin (bLg)

LGB; B-LG

Specificity

This assay has high sensitivity and excellent specificity for detection of Beta-Lactoglobulin (bLg).
No significant cross-reactivity or interference between Beta-Lactoglobulin (bLg) and analogues was observed.

Recovery

Matrices listed below were spiked with certain level of recombinant Beta-Lactoglobulin (bLg) and the recovery rates were calculated by comparing the measured value to the expected amount of Beta-Lactoglobulin (bLg) in samples.

Matrix Recovery range (%) Average(%)
serum(n=5) 92-105 99
EDTA plasma(n=5) 80-91 85
heparin plasma(n=5) 85-93 89

Precision

Intra-assay Precision (Precision within an assay): 3 samples with low, middle and high level Beta-Lactoglobulin (bLg) were tested 20 times on one plate, respectively.
Inter-assay Precision (Precision between assays): 3 samples with low, middle and high level Beta-Lactoglobulin (bLg) were tested on 3 different plates, 8 replicates in each plate.
CV(%) = SD/meanX100
Intra-Assay: CV<10%
Inter-Assay: CV<12%

Linearity

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of Beta-Lactoglobulin (bLg) and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.

Sample 1:2 1:4 1:8 1:16
serum(n=5) 93-102% 79-99% 78-97% 88-102%
EDTA plasma(n=5) 80-101% 83-102% 81-92% 96-103%
heparin plasma(n=5) 95-102% 80-89% 83-91% 83-103%

Stability

The stability of kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage condition.
To minimize extra influence on the performance, operation procedures and lab conditions, especially room temperature, air humidity, incubator temperature should be strictly controlled. It is also strongly suggested that the whole assay is performed by the same operator from the beginning to the end.

Reagents and materials provided

Reagents Quantity Reagents Quantity
Pre-coated, ready to use 96-well strip plate 1 Plate sealer for 96 wells 4
Standard 2 Standard Diluent 1×20mL
Detection Reagent A 1×120µL Assay Diluent A 1×12mL
Detection Reagent B 1×120µL Assay Diluent B 1×12mL
Substrate A 1×10mL Substrate B 1×2mL
Wash Buffer (30 × concentrate) 1×20mL Instruction manual 1

Assay procedure summary

1. Prepare all reagents, samples and standards;
2. Add 100µL standard or sample to each well. Incubate 1 hours at 37°C;
3. Aspirate and add 100µL prepared Detection Reagent A. Incubate 1 hour at 37°C;
4. Aspirate and wash 3 times;
5. Add 100µL prepared Detection Reagent B. Incubate 30 minutes at 37°C;
6. Aspirate and wash 5 times;
7. Add 100µL Substrate Solution. Incubate 10 minutes at 37°C;
8. Read RLU value immediately.

GIVEAWAYS

INCREMENT SERVICES

Magazine Citations
Journal of Dairy Science Proteomics and microstructure profiling of goat milk protein after homogenization Pubmed: 30827554
Catalog No. Related products for research use of Bos taurus; Bovine (Cattle) Organism species Applications (RESEARCH USE ONLY!)
NPB023Bo01 Native Beta-Lactoglobulin (bLg) Positive Control; Immunogen; SDS-PAGE; WB.
RPB023Bo01 Recombinant Beta-Lactoglobulin (bLg) Positive Control; Immunogen; SDS-PAGE; WB.
PAB023Bo01 Polyclonal Antibody to Beta-Lactoglobulin (bLg) WB; IHC; ICC; IP.
PAB023Bo05 Polyclonal Antibody to Beta-Lactoglobulin (bLg) WB; IHC; ICC; IP.
MAB023Bo21 Monoclonal Antibody to Beta-Lactoglobulin (bLg) WB; IHC; ICC; IP.
SEB023Bo ELISA Kit for Beta-Lactoglobulin (bLg) Enzyme-linked immunosorbent assay for Antigen Detection.
SCB023Bo CLIA Kit for Beta-Lactoglobulin (bLg) Chemiluminescent immunoassay for Antigen Detection.
LMB023Bo Multiplex Assay Kit for Beta-Lactoglobulin (bLg) ,etc. by FLIA (Flow Luminescence Immunoassay) FLIA Kit for Antigen Detection.
KSB023Bo01 ELISA Kit DIY Materials for Beta-Lactoglobulin (bLg) Main materials for "Do It (ELISA Kit) Yourself".