CLIA Kit for Platelet Derived Growth Factor Receptor Like Protein (PDGFRL)

PRLTS; PDGF-Receptor Beta-Like Tumor Suppressor

Specificity

This assay has high sensitivity and excellent specificity for detection of Platelet Derived Growth Factor Receptor Like Protein (PDGFRL).
No significant cross-reactivity or interference between Platelet Derived Growth Factor Receptor Like Protein (PDGFRL) and analogues was observed.

Recovery

Matrices listed below were spiked with certain level of recombinant Platelet Derived Growth Factor Receptor Like Protein (PDGFRL) and the recovery rates were calculated by comparing the measured value to the expected amount of Platelet Derived Growth Factor Receptor Like Protein (PDGFRL) in samples.

Matrix Recovery range (%) Average(%)
serum(n=5) 80-88 80
EDTA plasma(n=5) 78-101 82
heparin plasma(n=5) 81-103 89

Precision

Intra-assay Precision (Precision within an assay): 3 samples with low, middle and high level Platelet Derived Growth Factor Receptor Like Protein (PDGFRL) were tested 20 times on one plate, respectively.
Inter-assay Precision (Precision between assays): 3 samples with low, middle and high level Platelet Derived Growth Factor Receptor Like Protein (PDGFRL) were tested on 3 different plates, 8 replicates in each plate.
CV(%) = SD/meanX100
Intra-Assay: CV<10%
Inter-Assay: CV<12%

Linearity

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of Platelet Derived Growth Factor Receptor Like Protein (PDGFRL) and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.

Sample 1:2 1:4 1:8 1:16
serum(n=5) 97-105% 80-89% 92-105% 89-98%
EDTA plasma(n=5) 83-99% 85-92% 95-104% 90-105%
heparin plasma(n=5) 94-103% 95-103% 90-103% 84-95%

Stability

The stability of kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage condition.
To minimize extra influence on the performance, operation procedures and lab conditions, especially room temperature, air humidity, incubator temperature should be strictly controlled. It is also strongly suggested that the whole assay is performed by the same operator from the beginning to the end.

Reagents and materials provided

Reagents Quantity Reagents Quantity
Pre-coated, ready to use 96-well strip plate 1 Plate sealer for 96 wells 4
Standard 2 Standard Diluent 1×20mL
Detection Reagent A 1×120µL Assay Diluent A 1×12mL
Detection Reagent B 1×120µL Assay Diluent B 1×12mL
Substrate A 1×10mL Substrate B 1×2mL
Wash Buffer (30 × concentrate) 1×20mL Instruction manual 1

Assay procedure summary

1. Prepare all reagents, samples and standards;
2. Add 100µL standard or sample to each well. Incubate 1 hours at 37°C;
3. Aspirate and add 100µL prepared Detection Reagent A. Incubate 1 hour at 37°C;
4. Aspirate and wash 3 times;
5. Add 100µL prepared Detection Reagent B. Incubate 30 minutes at 37°C;
6. Aspirate and wash 5 times;
7. Add 100µL Substrate Solution. Incubate 10 minutes at 37°C;
8. Read RLU value immediately.

GIVEAWAYS

INCREMENT SERVICES

Catalog No. Related products for research use of Homo sapiens (Human) Organism species Applications (RESEARCH USE ONLY!)
RPC922Hu01 Recombinant Platelet Derived Growth Factor Receptor Like Protein (PDGFRL) Positive Control; Immunogen; SDS-PAGE; WB.
PAC922Hu01 Polyclonal Antibody to Platelet Derived Growth Factor Receptor Like Protein (PDGFRL) WB; IHC; ICC; IP.
MAC922Hu22 Monoclonal Antibody to Platelet Derived Growth Factor Receptor Like Protein (PDGFRL) WB; IHC; ICC; IP.
SEC922Hu ELISA Kit for Platelet Derived Growth Factor Receptor Like Protein (PDGFRL) Enzyme-linked immunosorbent assay for Antigen Detection.
SCC922Hu CLIA Kit for Platelet Derived Growth Factor Receptor Like Protein (PDGFRL) Chemiluminescent immunoassay for Antigen Detection.