ELISA Kit for Immunoglobulin G3 (IgG3)

IGHG3; HDC; Heavy chain disease protein; Ig Gamma-3 Chain C Region; Immunoglobulin Heavy Constant Gamma 3; G3m Marker; Immunoglobulin Gm3

Specificity

This assay has high sensitivity and excellent specificity for detection of Immunoglobulin G3 (IgG3).
No significant cross-reactivity or interference between Immunoglobulin G3 (IgG3) and analogues was observed.

Recovery

Matrices listed below were spiked with certain level of recombinant Immunoglobulin G3 (IgG3) and the recovery rates were calculated by comparing the measured value to the expected amount of Immunoglobulin G3 (IgG3) in samples.

Matrix Recovery range (%) Average(%)
serum(n=5) 96-105 101
EDTA plasma(n=5) 86-101 96
heparin plasma(n=5) 84-91 87

Precision

Intra-assay Precision (Precision within an assay): 3 samples with low, middle and high level Immunoglobulin G3 (IgG3) were tested 20 times on one plate, respectively.
Inter-assay Precision (Precision between assays): 3 samples with low, middle and high level Immunoglobulin G3 (IgG3) were tested on 3 different plates, 8 replicates in each plate.
CV(%) = SD/meanX100
Intra-Assay: CV<10%
Inter-Assay: CV<12%

Linearity

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of Immunoglobulin G3 (IgG3) and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.

Sample 1:2 1:4 1:8 1:16
serum(n=5) 97-104% 81-92% 97-105% 80-96%
EDTA plasma(n=5) 80-89% 85-97% 95-103% 98-105%
heparin plasma(n=5) 82-96% 87-95% 91-99% 79-101%

Stability

The stability of kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage condition.
To minimize extra influence on the performance, operation procedures and lab conditions, especially room temperature, air humidity, incubator temperature should be strictly controlled. It is also strongly suggested that the whole assay is performed by the same operator from the beginning to the end.

Reagents and materials provided

Reagents Quantity Reagents Quantity
Pre-coated, ready to use 96-well strip plate 1 Plate sealer for 96 wells 4
Standard 2 Standard Diluent 1×20mL
Detection Reagent A 1×120µL Assay Diluent A 1×12mL
Detection Reagent B 1×120µL Assay Diluent B 1×12mL
TMB Substrate 1×9mL Stop Solution 1×6mL
Wash Buffer (30 × concentrate) 1×20mL Instruction manual 1

Assay procedure summary

1. Prepare all reagents, samples and standards;
2. Add 100µL standard or sample to each well. Incubate 1 hours at 37°C;
3. Aspirate and add 100µL prepared Detection Reagent A. Incubate 1 hour at 37°C;
4. Aspirate and wash 3 times;
5. Add 100µL prepared Detection Reagent B. Incubate 30 minutes at 37°C;
6. Aspirate and wash 5 times;
7. Add 90µL Substrate Solution. Incubate 10-20 minutes at 37°C;
8. Add 50µL Stop Solution. Read at 450nm immediately.

GIVEAWAYS

INCREMENT SERVICES

Magazine Citations
BioMed Research International The Effect of CXC Motif Chemokine 13 on Hepatocellular Carcinoma Associates with Wnt Signaling Hindawi:Source
Oncology Reports Beneficial effect of T follicular helper cells on antibody class switching of B cells in prostate cancer Pubmed:25529861
Biomed Res Int The Effect of C-X-C Motif Chemokine 13 on Hepatocellular Carcinoma Associates with Wnt Signaling PubMed: 26161394
Catalog No. Related products for research use of Homo sapiens (Human) Organism species Applications (RESEARCH USE ONLY!)
NPA829Hu01 Native Immunoglobulin G3 (IgG3) Positive Control; Immunogen; SDS-PAGE; WB.
RPA829Hu01 Recombinant Immunoglobulin G3 (IgG3) Positive Control; Immunogen; SDS-PAGE; WB.
PAA829Hu01 Polyclonal Antibody to Immunoglobulin G3 (IgG3) WB; IHC; ICC; IP.
LAA829Hu81 FITC-Linked Polyclonal Antibody to Immunoglobulin G3 (IgG3) WB; IHC; ICC; IF.
LAA829Hu71 Biotin-Linked Polyclonal Antibody to Immunoglobulin G3 (IgG3) WB; IHC; ICC.
MAA829Hu22 Monoclonal Antibody to Immunoglobulin G3 (IgG3) WB; IHC; ICC; IP.
LAA829Hu72 Biotin-Linked Monoclonal Antibody to Immunoglobulin G3 (IgG3) WB; IHC; ICC.
LAA829Hu82 FITC-Linked Monoclonal Antibody to Immunoglobulin G3 (IgG3) WB; IHC; ICC; IF.
SEA829Hu ELISA Kit for Immunoglobulin G3 (IgG3) Enzyme-linked immunosorbent assay for Antigen Detection.
LMA829Hu Multiplex Assay Kit for Immunoglobulin G3 (IgG3) ,etc. by FLIA (Flow Luminescence Immunoassay) FLIA Kit for Antigen Detection.
KSA829Hu01 ELISA Kit DIY Materials for Immunoglobulin G3 (IgG3) Main materials for "Do It (ELISA Kit) Yourself".