Packages (Simulation)
Reagent Preparation
Image (I)
Image (II)
Certificate
ELISA Kit for Cluster Of Differentiation 8a (CD8a)
p32; CD8-A; Leu2; MAL; T-cell surface glycoprotein CD8 alpha chain; T-lymphocyte differentiation antigen T8/Leu-2
- Product No.SEB099Ga
- Organism SpeciesChicken (Gallus) Same name, Different species.
- Sample Typetissue homogenates, cell lysates and other biological fluids
- Test MethodDouble-antibody Sandwich
- Assay Length3h
- Detection Range0.156-10ng/mL
- SensitivityThe minimum detectable dose of this kit is typically less than 0.059ng/mL.
- DownloadInstruction Manual
- UOM 48T96T 96T*5 96T*10 96T*100
- FOB
US$ 505
For more details, please contact local distributors! US$ 722 US$ 3249 US$ 6137 US$ 50540
Specificity
This assay has high sensitivity and excellent specificity for detection of Cluster Of Differentiation 8a (CD8a).
No significant cross-reactivity or interference between Cluster Of Differentiation 8a (CD8a) and analogues was observed.
Precision
Intra-assay Precision (Precision within an assay): 3 samples with low, middle and high level Cluster Of Differentiation 8a (CD8a) were tested 20 times on one plate, respectively.
Inter-assay Precision (Precision between assays): 3 samples with low, middle and high level Cluster Of Differentiation 8a (CD8a) were tested on 3 different plates, 8 replicates in each plate.
CV(%) = SD/meanX100
Intra-Assay: CV<10%
Inter-Assay: CV<12%
Stability
The stability of kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage condition.
To minimize extra influence on the performance, operation procedures and lab conditions, especially room temperature, air humidity, incubator temperature should be strictly controlled. It is also strongly suggested that the whole assay is performed by the same operator from the beginning to the end.
Reagents and materials provided
Reagents | Quantity | Reagents | Quantity |
Pre-coated, ready to use 96-well strip plate | 1 | Plate sealer for 96 wells | 4 |
Standard | 2 | Standard Diluent | 1×20mL |
Detection Reagent A | 1×120µL | Assay Diluent A | 1×12mL |
Detection Reagent B | 1×120µL | Assay Diluent B | 1×12mL |
TMB Substrate | 1×9mL | Stop Solution | 1×6mL |
Wash Buffer (30 × concentrate) | 1×20mL | Instruction manual | 1 |
Assay procedure summary
1. Prepare all reagents, samples and standards;
2. Add 100µL standard or sample to each well. Incubate 1 hours at 37°C;
3. Aspirate and add 100µL prepared Detection Reagent A. Incubate 1 hour at 37°C;
4. Aspirate and wash 3 times;
5. Add 100µL prepared Detection Reagent B. Incubate 30 minutes at 37°C;
6. Aspirate and wash 5 times;
7. Add 90µL Substrate Solution. Incubate 10-20 minutes at 37°C;
8. Add 50µL Stop Solution. Read at 450nm immediately.
GIVEAWAYS
INCREMENT SERVICES
- Single-component Reagents of Assay Kit
- Lysis Buffer Specific for ELISA / CLIA
- Quality Control of Kit
- ELISA Kit Customized Service
- Disease Model Customized Service
- Serums Customized Service
- TGFB1 Activation Reagent
- Real Time PCR Experimental Service
- Streptavidin
- Fast blue Protein Stain solution
- Single-component Reagents of FLIA Kit
- Streptavidin-Agarose Beads
Magazine | Citations |
Taylor & Francis | Influences of Hyriopsis cumingii polysaccharides on mice immunosignaling molecules and T lymphocyte differentiation 09540105.2017.1306494 |
Catalog No. | Related products for research use of Chicken (Gallus) Organism species | Applications (RESEARCH USE ONLY!) |
SEB099Ga | ELISA Kit for Cluster Of Differentiation 8a (CD8a) | Enzyme-linked immunosorbent assay for Antigen Detection. |
LMB099Ga | Multiplex Assay Kit for Cluster Of Differentiation 8a (CD8a) ,etc. by FLIA (Flow Luminescence Immunoassay) | FLIA Kit for Antigen Detection. |