ELISA Kit for Alpha-Fetoprotein Lens Culinaris Agglutinin 3 (aFPL3)

AFP-L3

Specificity

This assay has high sensitivity and excellent specificity for detection of Alpha-Fetoprotein Lens Culinaris Agglutinin 3 (aFPL3).
No significant cross-reactivity or interference between Alpha-Fetoprotein Lens Culinaris Agglutinin 3 (aFPL3) and analogues was observed.

Recovery

Matrices listed below were spiked with certain level of recombinant Alpha-Fetoprotein Lens Culinaris Agglutinin 3 (aFPL3) and the recovery rates were calculated by comparing the measured value to the expected amount of Alpha-Fetoprotein Lens Culinaris Agglutinin 3 (aFPL3) in samples.

Matrix Recovery range (%) Average(%)
serum(n=5) 81-103 99
EDTA plasma(n=5) 92-102 96
heparin plasma(n=5) 98-105 102

Precision

Intra-assay Precision (Precision within an assay): 3 samples with low, middle and high level Alpha-Fetoprotein Lens Culinaris Agglutinin 3 (aFPL3) were tested 20 times on one plate, respectively.
Inter-assay Precision (Precision between assays): 3 samples with low, middle and high level Alpha-Fetoprotein Lens Culinaris Agglutinin 3 (aFPL3) were tested on 3 different plates, 8 replicates in each plate.
CV(%) = SD/meanX100
Intra-Assay: CV<10%
Inter-Assay: CV<12%

Linearity

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of Alpha-Fetoprotein Lens Culinaris Agglutinin 3 (aFPL3) and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.

Sample 1:2 1:4 1:8 1:16
serum(n=5) 92-101% 90-97% 91-101% 81-103%
EDTA plasma(n=5) 87-94% 87-101% 87-103% 86-101%
heparin plasma(n=5) 90-97% 89-103% 80-96% 96-104%

Stability

The stability of kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage condition.
To minimize extra influence on the performance, operation procedures and lab conditions, especially room temperature, air humidity, incubator temperature should be strictly controlled. It is also strongly suggested that the whole assay is performed by the same operator from the beginning to the end.

Reagents and materials provided

Reagents Quantity Reagents Quantity
Pre-coated, ready to use 96-well strip plate 1 Plate sealer for 96 wells 4
Standard 2 Standard Diluent 1×20mL
Detection Reagent A 1×120µL Assay Diluent A 1×12mL
Detection Reagent B 1×120µL Assay Diluent B 1×12mL
TMB Substrate 1×9mL Stop Solution 1×6mL
Wash Buffer (30 × concentrate) 1×20mL Instruction manual 1

Assay procedure summary

1. Prepare all reagents, samples and standards;
2. Add 100µL standard or sample to each well. Incubate 1 hours at 37°C;
3. Aspirate and add 100µL prepared Detection Reagent A. Incubate 1 hour at 37°C;
4. Aspirate and wash 3 times;
5. Add 100µL prepared Detection Reagent B. Incubate 30 minutes at 37°C;
6. Aspirate and wash 5 times;
7. Add 90µL Substrate Solution. Incubate 10-20 minutes at 37°C;
8. Add 50µL Stop Solution. Read at 450nm immediately.

GIVEAWAYS

INCREMENT SERVICES

Magazine Citations
Oxid Antioxid Med Sci Oxidative stress markers, alpha-fetoprotein and alpha-fetoprotein-L3 levels of alcohol dependent subjects Ejmanager: Source
PLOS ONE Development of a Highly Sensitive Glycan Microarray for Quantifying AFP-L3 for Early Prediction of Hepatitis B Virus–Related Hepatocellular Carcinoma Pubmed:24927126
ADVANCED RESEARCHES Search for Effective Serum Tumor Markers for Early Diagnosis of Hepatocellular Carcinoma Associated with Hepatitis C
Catalog No. Related products for research use of Homo sapiens (Human) Organism species Applications (RESEARCH USE ONLY!)
SEB117Hu ELISA Kit for Alpha-Fetoprotein Lens Culinaris Agglutinin 3 (aFPL3) Enzyme-linked immunosorbent assay for Antigen Detection.
LMB117Hu Multiplex Assay Kit for Alpha-Fetoprotein Lens Culinaris Agglutinin 3 (aFPL3) ,etc. by FLIA (Flow Luminescence Immunoassay) FLIA Kit for Antigen Detection.