ELISA Kit for Monoamine Oxidase A (MAOA)

MAO-A; Amine oxidase [flavin-containing] A

Specificity

This assay has high sensitivity and excellent specificity for detection of Monoamine Oxidase A (MAOA).
No significant cross-reactivity or interference between Monoamine Oxidase A (MAOA) and analogues was observed.

Recovery

Matrices listed below were spiked with certain level of recombinant Monoamine Oxidase A (MAOA) and the recovery rates were calculated by comparing the measured value to the expected amount of Monoamine Oxidase A (MAOA) in samples.

Matrix Recovery range (%) Average(%)
serum(n=5) 83-97 89
EDTA plasma(n=5) 83-103 92
heparin plasma(n=5) 81-95 87

Precision

Intra-assay Precision (Precision within an assay): 3 samples with low, middle and high level Monoamine Oxidase A (MAOA) were tested 20 times on one plate, respectively.
Inter-assay Precision (Precision between assays): 3 samples with low, middle and high level Monoamine Oxidase A (MAOA) were tested on 3 different plates, 8 replicates in each plate.
CV(%) = SD/meanX100
Intra-Assay: CV<10%
Inter-Assay: CV<12%

Linearity

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of Monoamine Oxidase A (MAOA) and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.

Sample 1:2 1:4 1:8 1:16
serum(n=5) 85-94% 84-92% 99-105% 92-105%
EDTA plasma(n=5) 81-105% 89-102% 94-101% 84-95%
heparin plasma(n=5) 94-102% 78-95% 84-92% 88-104%

Stability

The stability of kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage condition.
To minimize extra influence on the performance, operation procedures and lab conditions, especially room temperature, air humidity, incubator temperature should be strictly controlled. It is also strongly suggested that the whole assay is performed by the same operator from the beginning to the end.

Reagents and materials provided

Reagents Quantity Reagents Quantity
Pre-coated, ready to use 96-well strip plate 1 Plate sealer for 96 wells 4
Standard 2 Standard Diluent 1×20mL
Detection Reagent A 1×120µL Assay Diluent A 1×12mL
Detection Reagent B 1×120µL Assay Diluent B 1×12mL
TMB Substrate 1×9mL Stop Solution 1×6mL
Wash Buffer (30 × concentrate) 1×20mL Instruction manual 1

Assay procedure summary

1. Prepare all reagents, samples and standards;
2. Add 100µL standard or sample to each well. Incubate 1 hours at 37°C;
3. Aspirate and add 100µL prepared Detection Reagent A. Incubate 1 hour at 37°C;
4. Aspirate and wash 3 times;
5. Add 100µL prepared Detection Reagent B. Incubate 30 minutes at 37°C;
6. Aspirate and wash 5 times;
7. Add 90µL Substrate Solution. Incubate 10-20 minutes at 37°C;
8. Add 50µL Stop Solution. Read at 450nm immediately.

GIVEAWAYS

INCREMENT SERVICES

Magazine Citations
brain research Ellagic acid exerts protective effect in intrastriatal 6-hydroxydopamine rat model of Parkinson's disease: Possible involvement of ERβ/Nrf2/HO-1 signaling. pubmed:28238669
PLoS One Neuroprotective effect of sodium alginate against chromium-induced brain damage in rats Pubmed:35421180
Catalog No. Related products for research use of Mus musculus (Mouse) Organism species Applications (RESEARCH USE ONLY!)
RPB954Mu02 Recombinant Monoamine Oxidase A (MAOA) Positive Control; Immunogen; SDS-PAGE; WB.
RPB954Mu01 Recombinant Monoamine Oxidase A (MAOA) Positive Control; Immunogen; SDS-PAGE; WB.
PAB954Mu02 Polyclonal Antibody to Monoamine Oxidase A (MAOA) WB; IHC; ICC; IP.
PAB954Mu01 Polyclonal Antibody to Monoamine Oxidase A (MAOA) WB; IHC; ICC; IP.
LAB954Mu71 Biotin-Linked Polyclonal Antibody to Monoamine Oxidase A (MAOA) WB; IHC; ICC.
SEB954Mu ELISA Kit for Monoamine Oxidase A (MAOA) Enzyme-linked immunosorbent assay for Antigen Detection.
LMB954Mu Multiplex Assay Kit for Monoamine Oxidase A (MAOA) ,etc. by FLIA (Flow Luminescence Immunoassay) FLIA Kit for Antigen Detection.