ELISA Kit for Leukocyte Cell Derived Chemotaxin 2 (LECT2)

Chm-II; Chm2

Specificity

This assay has high sensitivity and excellent specificity for detection of Leukocyte Cell Derived Chemotaxin 2 (LECT2).
No significant cross-reactivity or interference between Leukocyte Cell Derived Chemotaxin 2 (LECT2) and analogues was observed.

Recovery

Matrices listed below were spiked with certain level of recombinant Leukocyte Cell Derived Chemotaxin 2 (LECT2) and the recovery rates were calculated by comparing the measured value to the expected amount of Leukocyte Cell Derived Chemotaxin 2 (LECT2) in samples.

Matrix Recovery range (%) Average(%)
serum(n=5) 82-103 93
EDTA plasma(n=5) 96-105 101
heparin plasma(n=5) 78-90 82

Precision

Intra-assay Precision (Precision within an assay): 3 samples with low, middle and high level Leukocyte Cell Derived Chemotaxin 2 (LECT2) were tested 20 times on one plate, respectively.
Inter-assay Precision (Precision between assays): 3 samples with low, middle and high level Leukocyte Cell Derived Chemotaxin 2 (LECT2) were tested on 3 different plates, 8 replicates in each plate.
CV(%) = SD/meanX100
Intra-Assay: CV<10%
Inter-Assay: CV<12%

Linearity

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of Leukocyte Cell Derived Chemotaxin 2 (LECT2) and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.

Sample 1:2 1:4 1:8 1:16
serum(n=5) 99-105% 86-93% 78-99% 78-91%
EDTA plasma(n=5) 88-102% 93-105% 88-95% 95-102%
heparin plasma(n=5) 96-105% 80-90% 80-96% 80-94%

Stability

The stability of kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage condition.
To minimize extra influence on the performance, operation procedures and lab conditions, especially room temperature, air humidity, incubator temperature should be strictly controlled. It is also strongly suggested that the whole assay is performed by the same operator from the beginning to the end.

Reagents and materials provided

Reagents Quantity Reagents Quantity
Pre-coated, ready to use 96-well strip plate 1 Plate sealer for 96 wells 4
Standard 2 Standard Diluent 1×20mL
Detection Reagent A 1×120µL Assay Diluent A 1×12mL
Detection Reagent B 1×120µL Assay Diluent B 1×12mL
TMB Substrate 1×9mL Stop Solution 1×6mL
Wash Buffer (30 × concentrate) 1×20mL Instruction manual 1

Assay procedure summary

1. Prepare all reagents, samples and standards;
2. Add 100µL standard or sample to each well. Incubate 1 hours at 37°C;
3. Aspirate and add 100µL prepared Detection Reagent A. Incubate 1 hour at 37°C;
4. Aspirate and wash 3 times;
5. Add 100µL prepared Detection Reagent B. Incubate 30 minutes at 37°C;
6. Aspirate and wash 5 times;
7. Add 90µL Substrate Solution. Incubate 10-20 minutes at 37°C;
8. Add 50µL Stop Solution. Read at 450nm immediately.

GIVEAWAYS

INCREMENT SERVICES

Magazine Citations
Translational Gastroenterology and Hepatology Leukocyte cell derived chemotaxin-2 (Lect2) as a predictor of survival in adult acute liver failure Pubmed: 30976720
J Endocrinol Leukocyte cell-derived chemotaxin 2 affects nonalcoholic fatty liver disease Pubmed: 32485675
肝脏 血清LECT2 检测对肝细胞癌诊断的意义
Catalog No. Related products for research use of Homo sapiens (Human) Organism species Applications (RESEARCH USE ONLY!)
EPF541Hu61 Eukaryotic Leukocyte Cell Derived Chemotaxin 2 (LECT2) Positive Control; Immunogen; SDS-PAGE; WB.
APF541Hu61 Active Leukocyte Cell Derived Chemotaxin 2 (LECT2) Cell culture; Activity Assays.
RPF541Hu01 Recombinant Leukocyte Cell Derived Chemotaxin 2 (LECT2) Positive Control; Immunogen; SDS-PAGE; WB.
PAF541Hu01 Polyclonal Antibody to Leukocyte Cell Derived Chemotaxin 2 (LECT2) WB; IHC; ICC; IP.
MAF541Hu21 Monoclonal Antibody to Leukocyte Cell Derived Chemotaxin 2 (LECT2) WB; IHC; ICC; IP.
SEF541Hu ELISA Kit for Leukocyte Cell Derived Chemotaxin 2 (LECT2) Enzyme-linked immunosorbent assay for Antigen Detection.
LMF541Hu Multiplex Assay Kit for Leukocyte Cell Derived Chemotaxin 2 (LECT2) ,etc. by FLIA (Flow Luminescence Immunoassay) FLIA Kit for Antigen Detection.