ELISA Kit for CD200 Receptor 2 (CD200R2)

CD200R1L; CD200RLa; HuCD200R2; Cell surface glycoprotein CD200 receptor 2; CD200 cell surface glycoprotein receptor-like a; Cell surface glycoprotein OX2 receptor 2

Specificity

This assay has high sensitivity and excellent specificity for detection of CD200 Receptor 2 (CD200R2).
No significant cross-reactivity or interference between CD200 Receptor 2 (CD200R2) and analogues was observed.

Recovery

Matrices listed below were spiked with certain level of recombinant CD200 Receptor 2 (CD200R2) and the recovery rates were calculated by comparing the measured value to the expected amount of CD200 Receptor 2 (CD200R2) in samples.

Matrix Recovery range (%) Average(%)
serum(n=5) 87-101 95
EDTA plasma(n=5) 78-101 92
heparin plasma(n=5) 79-95 82

Precision

Intra-assay Precision (Precision within an assay): 3 samples with low, middle and high level CD200 Receptor 2 (CD200R2) were tested 20 times on one plate, respectively.
Inter-assay Precision (Precision between assays): 3 samples with low, middle and high level CD200 Receptor 2 (CD200R2) were tested on 3 different plates, 8 replicates in each plate.
CV(%) = SD/meanX100
Intra-Assay: CV<10%
Inter-Assay: CV<12%

Linearity

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of CD200 Receptor 2 (CD200R2) and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.

Sample 1:2 1:4 1:8 1:16
serum(n=5) 93-101% 98-105% 94-101% 93-105%
EDTA plasma(n=5) 79-95% 86-93% 88-95% 91-98%
heparin plasma(n=5) 93-101% 87-94% 80-97% 89-101%

Stability

The stability of kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage condition.
To minimize extra influence on the performance, operation procedures and lab conditions, especially room temperature, air humidity, incubator temperature should be strictly controlled. It is also strongly suggested that the whole assay is performed by the same operator from the beginning to the end.

Reagents and materials provided

Reagents Quantity Reagents Quantity
Pre-coated, ready to use 96-well strip plate 1 Plate sealer for 96 wells 4
Standard 2 Standard Diluent 1×20mL
Detection Reagent A 1×120µL Assay Diluent A 1×12mL
Detection Reagent B 1×120µL Assay Diluent B 1×12mL
TMB Substrate 1×9mL Stop Solution 1×6mL
Wash Buffer (30 × concentrate) 1×20mL Instruction manual 1

Assay procedure summary

1. Prepare all reagents, samples and standards;
2. Add 100µL standard or sample to each well. Incubate 1 hours at 37°C;
3. Aspirate and add 100µL prepared Detection Reagent A. Incubate 1 hour at 37°C;
4. Aspirate and wash 3 times;
5. Add 100µL prepared Detection Reagent B. Incubate 30 minutes at 37°C;
6. Aspirate and wash 5 times;
7. Add 90µL Substrate Solution. Incubate 10-20 minutes at 37°C;
8. Add 50µL Stop Solution. Read at 450nm immediately.

GIVEAWAYS

INCREMENT SERVICES

Catalog No. Related products for research use of Homo sapiens (Human) Organism species Applications (RESEARCH USE ONLY!)
RPT653Hu01 Recombinant CD200 Receptor 2 (CD200R2) Positive Control; Immunogen; SDS-PAGE; WB.
PAT653Hu01 Polyclonal Antibody to CD200 Receptor 2 (CD200R2) WB; IHC; ICC; IP.
MAT653Hu21 Monoclonal Antibody to CD200 Receptor 2 (CD200R2) WB; IHC; ICC; IP.
SET653Hu ELISA Kit for CD200 Receptor 2 (CD200R2) Enzyme-linked immunosorbent assay for Antigen Detection.
LMT653Hu Multiplex Assay Kit for CD200 Receptor 2 (CD200R2) ,etc. by FLIA (Flow Luminescence Immunoassay) FLIA Kit for Antigen Detection.