ELISA Kit for Cholesterol (CH)

Specificity

This assay has high sensitivity and excellent specificity for detection of Cholesterol (CH).
No significant cross-reactivity or interference between Cholesterol (CH) and analogues was observed.

Recovery

Matrices listed below were spiked with certain level of Cholesterol (CH) and the recovery rates were calculated by comparing the measured value to the expected amount of Cholesterol (CH) in samples.

Matrix Recovery range (%) Average(%)
serum(n=5) 81-102 86
EDTA plasma(n=5) 84-103 89
heparin plasma(n=5) 80-97 93

Precision

Intra-assay Precision (Precision within an assay): 3 samples with low, middle and high level Cholesterol (CH) were tested 20 times on one plate, respectively.
Inter-assay Precision (Precision between assays): 3 samples with low, middle and high level Cholesterol (CH) were tested on 3 different plates, 8 replicates in each plate.
CV(%) = SD/meanX100
Intra-Assay: CV<10%
Inter-Assay: CV<12%

Linearity

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of Cholesterol (CH) and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.

Sample 1:2 1:4 1:8 1:16
serum(n=5) 95-103% 80-93% 87-94% 94-104%
EDTA plasma(n=5) 81-91% 78-89% 88-105% 97-104%
heparin plasma(n=5) 80-102% 93-102% 79-94% 86-105%

Stability

The stability of kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage condition.
To minimize extra influence on the performance, operation procedures and lab conditions, especially room temperature, air humidity, incubator temperature should be strictly controlled. It is also strongly suggested that the whole assay is performed by the same operator from the beginning to the end.

Reagents and materials provided

Reagents Quantity Reagents Quantity
Pre-coated, ready to use 96-well strip plate 1 Plate sealer for 96 wells 4
Standard 2 Standard Diluent 1×20mL
Detection Reagent A 1×120µL Assay Diluent A 1×12mL
Detection Reagent B 1×120µL Assay Diluent B 1×12mL
TMB Substrate 1×9mL Stop Solution 1×6mL
Wash Buffer (30 × concentrate) 1×20mL Instruction manual 1

Assay procedure summary

1. Prepare all reagents, samples and standards;
2. Add 50µL standard or sample to each well.
    And then add 50µL prepared Detection Reagent A immediately.
    Shake and mix. Incubate 1 hour at 37°C;
3. Aspirate and wash 3 times;
4. Add 100µL prepared Detection Reagent B. Incubate 30 minutes at 37°C;
5. Aspirate and wash 5 times;
6. Add 90µL Substrate Solution. Incubate 10-20 minutes at 37°C;
7. Add 50µL Stop Solution. Read at 450 nm immediately.

GIVEAWAYS

INCREMENT SERVICES

Magazine Citations
Journal of Cell Science Nuclear phosphatidylinositol 4, 5-bisphosphate islets contribute to efficient RNA polymerase II-dependent transcription Pubmed:29507116
Journal of Diabetes Research Regulation and Mechanism of miR-518d through the PPARα-Mediated NF-κB Pathway in the Development of Gestational Diabetes Mellitus Pubmed: 33123597
Vascul Pharmacol Immune responses against oxidized LDL as possible targets for prevention of atherosclerosis in systemic lupus erythematosus 33857652
Evid Based Complement Alternat Med Effect of Acupoint Embedding on Serum Leptin and Hypothalamus Leptin Receptor Expression in Rats with Simple Obesity 34659432
Nat Commun GP73 is a TBC-domain Rab GTPase-activating protein contributing to the pathogenesis of non-alcoholic fatty liver disease without obesity 34853313
Antioxidants Antioxidant Effect of Tyr-Ala Extracted from Zein on INS-1 Cells and Type 2 Diabetes High-Fat-Diet-Induced Mice Pubmed:35740008
Catalog No. Related products for research use of Pan-species (General) Organism species Applications (RESEARCH USE ONLY!)
CPB701Ge11 BSA Conjugated Cholesterol (CH) Immunogen; SDS-PAGE; WB.
CPB701Ge21 OVA Conjugated Cholesterol (CH) Immunogen; SDS-PAGE; WB.
PAB701Ge01 Polyclonal Antibody to Cholesterol (CH) ELISA, CLIA. / IHC-Fr, ICC, IP (predicted).
LAB701Ge71 Biotin-Linked Polyclonal Antibody to Cholesterol (CH) IHC; ICC; IP.
CEB701Ge ELISA Kit for Cholesterol (CH) Enzyme-linked immunosorbent assay for Antigen Detection.
LMB701Ge Multiplex Assay Kit for Cholesterol (CH) ,etc. by FLIA (Flow Luminescence Immunoassay) FLIA Kit for Antigen Detection.
KSB701Ge11 ELISA Kit DIY Materials for Cholesterol (CH) Main materials for "Do It (ELISA Kit) Yourself".