Multiplex Assay Kit for HLA Class II Histocompatibility Antigen, DRB1 Beta Chain (HLA-DRB1) ,etc. by FLIA (Flow Luminescence Immunoassay)

HLA-DRB1; MHCDRb1; HLADRb1; Major Histocompatibility Complex Class II DR Beta 1; HLA Class II Histocompatibility Antigen,DRB1-9 Beta Chain; HLA class II histocompatibility antigen, DRB1-15 beta chain

(Note: Up to 8-plex in one testing reaction)

Specificity

This assay has high sensitivity and excellent specificity for detection of HLA Class II Histocompatibility Antigen, DRB1 Beta Chain (HLA-DRB1) ,etc. by FLIA (Flow Luminescence Immunoassay).
No significant cross-reactivity or interference between HLA Class II Histocompatibility Antigen, DRB1 Beta Chain (HLA-DRB1) ,etc. by FLIA (Flow Luminescence Immunoassay) and analogues was observed.

Recovery

Matrices listed below were spiked with certain level of recombinant HLA Class II Histocompatibility Antigen, DRB1 Beta Chain (HLA-DRB1) ,etc. by FLIA (Flow Luminescence Immunoassay) and the recovery rates were calculated by comparing the measured value to the expected amount of HLA Class II Histocompatibility Antigen, DRB1 Beta Chain (HLA-DRB1) ,etc. by FLIA (Flow Luminescence Immunoassay) in samples.

Matrix Recovery range (%) Average(%)
serum(n=5) 99-105 102
EDTA plasma(n=5) 80-96 92
heparin plasma(n=5) 85-93 90
sodium citrate plasma(n=5) 89-99 96

Precision

Intra-assay Precision (Precision within an assay): 3 samples with low, middle and high level HLA Class II Histocompatibility Antigen, DRB1 Beta Chain (HLA-DRB1) ,etc. by FLIA (Flow Luminescence Immunoassay) were tested 20 times on one plate, respectively.
Inter-assay Precision (Precision between assays): 3 samples with low, middle and high level HLA Class II Histocompatibility Antigen, DRB1 Beta Chain (HLA-DRB1) ,etc. by FLIA (Flow Luminescence Immunoassay) were tested on 3 different plates, 8 replicates in each plate.
CV(%) = SD/meanX100
Intra-Assay: CV<10%
Inter-Assay: CV<12%

Linearity

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of HLA Class II Histocompatibility Antigen, DRB1 Beta Chain (HLA-DRB1) ,etc. by FLIA (Flow Luminescence Immunoassay) and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.

Sample 1:2 1:4 1:8 1:16
serum(n=5) 89-96% 92-99% 85-94% 81-98%
EDTA plasma(n=5) 93-101% 85-92% 78-101% 95-102%
heparin plasma(n=5) 89-96% 80-91% 78-98% 85-94%
sodium citrate plasma(n=5) 98-105% 89-102% 96-105% 78-94%

Stability

The stability of kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage condition.
To minimize extra influence on the performance, operation procedures and lab conditions, especially room temperature, air humidity, incubator temperature should be strictly controlled. It is also strongly suggested that the whole assay is performed by the same operator from the beginning to the end.

Reagents and materials provided

Reagents Quantity Reagents Quantity
96-well plate 1 Plate sealer for 96 wells 4
Pre-Mixed Standard 2 Standard Diluent 1×20mL
Pre-Mixed Magnetic beads (22#:HLA-DRB1) 1 Analysis buffer 1×20mL
Pre-Mixed Detection Reagent A 1×120μL Assay Diluent A 1×12mL
Detection Reagent B (PE-SA) 1×120μL Assay Diluent B 1×12mL
Sheath Fluid 1×10mL Wash Buffer (30 × concentrate) 1×20mL
Instruction manual 1

Assay procedure summary

1. Preparation of standards, reagents and samples before the experiment;
2. Add 100μL standard or sample to each well,
    add 10μL magnetic beads, and incubate 90min at 37°C on shaker;
3. Remove liquid on magnetic frame, add 100μL prepared Detection Reagent A. Incubate 60min at 37°C on shaker;
4. Wash plate on magnetic frame for three times;
5. Add 100μL prepared Detection Reagent B, and incubate 30 min at 37°C on shaker;
6. Wash plate on magnetic frame for three times;
7. Add 100μL sheath solution, swirl for 2 minutes, read on the machine.

GIVEAWAYS

INCREMENT SERVICES

Catalog No. Related products for research use of Homo sapiens (Human) Organism species Applications (RESEARCH USE ONLY!)
RPC167Hu01 Recombinant HLA Class II Histocompatibility Antigen, DRB1 Beta Chain (HLA-DRB1) Positive Control; Immunogen; SDS-PAGE; WB.
PAC167Hu01 Polyclonal Antibody to HLA Class II Histocompatibility Antigen, DRB1 Beta Chain (HLA-DRB1) WB; IHC
FAC167Hu01 Anti-HLA Class II Histocompatibility Antigen, DRB1 Beta Chain (HLA-DRB1) Polyclonal Antibody Flow cytometry.
FAC167Hu41 PE-Linked Anti-HLA Class II Histocompatibility Antigen, DRB1 Beta Chain (HLA-DRB1) Polyclonal Antibody Flow cytometry.
FAC167Hu51 APC-Linked Anti-HLA Class II Histocompatibility Antigen, DRB1 Beta Chain (HLA-DRB1) Polyclonal Antibody Flow cytometry.
FAC167Hu81 FITC-Linked Anti-HLA Class II Histocompatibility Antigen, DRB1 Beta Chain (HLA-DRB1) Polyclonal Antibody Flow cytometry.
MAC167Hu21 Monoclonal Antibody to HLA Class II Histocompatibility Antigen, DRB1 Beta Chain (HLA-DRB1) WB; IHC
MAC167Hu24 Monoclonal Antibody to HLA Class II Histocompatibility Antigen, DRB1 Beta Chain (HLA-DRB1) WB; IHC; ICC; IP.
MAC167Hu25 Monoclonal Antibody to HLA Class II Histocompatibility Antigen, DRB1 Beta Chain (HLA-DRB1) WB; IHC; ICC; IP.
MAC167Hu23 Monoclonal Antibody to HLA Class II Histocompatibility Antigen, DRB1 Beta Chain (HLA-DRB1) WB; IHC; ICC; IP.
MAC167Hu22 Monoclonal Antibody to HLA Class II Histocompatibility Antigen, DRB1 Beta Chain (HLA-DRB1) WB
SEC167Hu ELISA Kit for HLA Class II Histocompatibility Antigen, DRB1 Beta Chain (HLA-DRB1) Enzyme-linked immunosorbent assay for Antigen Detection.
LMC167Hu Multiplex Assay Kit for HLA Class II Histocompatibility Antigen, DRB1 Beta Chain (HLA-DRB1) ,etc. by FLIA (Flow Luminescence Immunoassay) FLIA Kit for Antigen Detection.