Mini Samples ELISA Kit for Pancreatic Polypeptide (PP)

PPY; PNP; Pancreatic Polypeptide Y; Obinepitide; Pancreatic icosapeptide; Pancreatic prohormone

Specificity

This assay has high sensitivity and excellent specificity for detection of Mini Samples Pancreatic Polypeptide (PP).
No significant cross-reactivity or interference between Mini Samples Pancreatic Polypeptide (PP) and analogues was observed.

Recovery

Matrices listed below were spiked with certain level of recombinant Mini Samples Pancreatic Polypeptide (PP) and the recovery rates were calculated by comparing the measured value to the expected amount of Mini Samples Pancreatic Polypeptide (PP) in samples.

Matrix Recovery range (%) Average(%)
serum(n=5) 96-104 99
EDTA plasma(n=5) 98-105 102
heparin plasma(n=5) 90-99 93

Precision

Intra-assay Precision (Precision within an assay): 3 samples with low, middle and high level Mini Samples Pancreatic Polypeptide (PP) were tested 20 times on one plate, respectively.
Inter-assay Precision (Precision between assays): 3 samples with low, middle and high level Mini Samples Pancreatic Polypeptide (PP) were tested on 3 different plates, 8 replicates in each plate.
CV(%) = SD/meanX100
Intra-Assay: CV<10%
Inter-Assay: CV<12%

Linearity

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of Mini Samples Pancreatic Polypeptide (PP) and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.

Sample 1:2 1:4 1:8 1:16
serum(n=5) 98-105% 79-105% 88-97% 85-97%
EDTA plasma(n=5) 80-97% 96-103% 81-97% 80-90%
heparin plasma(n=5) 86-97% 78-92% 89-97% 86-95%

Stability

The stability of kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage condition.
To minimize extra influence on the performance, operation procedures and lab conditions, especially room temperature, air humidity, incubator temperature should be strictly controlled. It is also strongly suggested that the whole assay is performed by the same operator from the beginning to the end.

Reagents and materials provided

Reagents Quantity Reagents Quantity
Pre-coated, ready to use 96-well strip plate 1 Plate sealer for 96 wells 4
Standard 2 Standard Diluent 1×20mL
Detection Reagent A 1×60µL Assay Diluent A 1×6mL
Detection Reagent B 1×60µL Assay Diluent B 1×6mL
TMB Substrate 1×9mL Stop Solution 1×3mL
Wash Buffer (30 × concentrate) 1×10mL Instruction manual 1

Assay procedure summary

1. Prepare all reagents, samples and standards;
2. Add 25µL standard or sample to each well.
    And then add 25μL prepared Detection Reagent A immediately.
    Shake and mix. Incubate 1 hour at 37°C;
3. Aspirate and wash 3 times;
4. Add 50µL prepared Detection Reagent B. Incubate 30 minutes at 37°C;
5. Aspirate and wash 5 times;
6. Add 50µL Substrate Solution. Incubate 10-20 minutes at 37°C;
7. Add 25µL Stop Solution. Read at 450 nm immediately.

GIVEAWAYS

INCREMENT SERVICES

Catalog No. Related products for research use of Rattus norvegicus (Rat) Organism species Applications (RESEARCH USE ONLY!)
RPB265Ra01 Recombinant Pancreatic Polypeptide (PP) Positive Control; Immunogen; SDS-PAGE; WB.
PAB265Ra01 Polyclonal Antibody to Pancreatic Polypeptide (PP) WB; IHC; ICC; IP.
MAB265Ra21 Monoclonal Antibody to Pancreatic Polypeptide (PP) WB; IHC; ICC; IP.
SEB265Ra ELISA Kit for Pancreatic Polypeptide (PP) Enzyme-linked immunosorbent assay for Antigen Detection.
CEB265Ra ELISA Kit for Pancreatic Polypeptide (PP) Enzyme-linked immunosorbent assay for Antigen Detection.
MEB265Ra Mini Samples ELISA Kit for Pancreatic Polypeptide (PP) Enzyme-linked immunosorbent assay for Antigen Detection.
LMB265Ra Multiplex Assay Kit for Pancreatic Polypeptide (PP) ,etc. by FLIA (Flow Luminescence Immunoassay) FLIA Kit for Antigen Detection.