ELISA Kit for Endothelin Converting Enzyme 1 (ECE1)

Specificity

This assay has high sensitivity and excellent specificity for detection of Endothelin Converting Enzyme 1 (ECE1).
No significant cross-reactivity or interference between Endothelin Converting Enzyme 1 (ECE1) and analogues was observed.

Recovery

Matrices listed below were spiked with certain level of recombinant Endothelin Converting Enzyme 1 (ECE1) and the recovery rates were calculated by comparing the measured value to the expected amount of Endothelin Converting Enzyme 1 (ECE1) in samples.

Matrix Recovery range (%) Average(%)
serum(n=5) 79-93 79
EDTA plasma(n=5) 97-105 101
heparin plasma(n=5) 91-98 94

Precision

Intra-assay Precision (Precision within an assay): 3 samples with low, middle and high level Endothelin Converting Enzyme 1 (ECE1) were tested 20 times on one plate, respectively.
Inter-assay Precision (Precision between assays): 3 samples with low, middle and high level Endothelin Converting Enzyme 1 (ECE1) were tested on 3 different plates, 8 replicates in each plate.
CV(%) = SD/meanX100
Intra-Assay: CV<10%
Inter-Assay: CV<12%

Linearity

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of Endothelin Converting Enzyme 1 (ECE1) and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.

Sample 1:2 1:4 1:8 1:16
serum(n=5) 93-101% 81-97% 78-95% 99-105%
EDTA plasma(n=5) 94-103% 78-101% 98-105% 92-101%
heparin plasma(n=5) 78-93% 89-102% 91-101% 89-101%

Stability

The stability of kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage condition.
To minimize extra influence on the performance, operation procedures and lab conditions, especially room temperature, air humidity, incubator temperature should be strictly controlled. It is also strongly suggested that the whole assay is performed by the same operator from the beginning to the end.

Reagents and materials provided

Reagents Quantity Reagents Quantity
Pre-coated, ready to use 96-well strip plate 1 Plate sealer for 96 wells 4
Standard 2 Standard Diluent 1×20mL
Detection Reagent A 1×120µL Assay Diluent A 1×12mL
Detection Reagent B 1×120µL Assay Diluent B 1×12mL
TMB Substrate 1×9mL Stop Solution 1×6mL
Wash Buffer (30 × concentrate) 1×20mL Instruction manual 1

Assay procedure summary

1. Prepare all reagents, samples and standards;
2. Add 100µL standard or sample to each well. Incubate 1 hours at 37°C;
3. Aspirate and add 100µL prepared Detection Reagent A. Incubate 1 hour at 37°C;
4. Aspirate and wash 3 times;
5. Add 100µL prepared Detection Reagent B. Incubate 30 minutes at 37°C;
6. Aspirate and wash 5 times;
7. Add 90µL Substrate Solution. Incubate 10-20 minutes at 37°C;
8. Add 50µL Stop Solution. Read at 450nm immediately.

GIVEAWAYS

INCREMENT SERVICES

Magazine Citations
Placenta Carbon monoxide attenuates bacteria-induced Endothelin-1 expression in second trimester placental explants Pubmed: 24731730
The Journal of Heart and Lung Transplantation The Role of the Endothelin-1 Pathway as a Biomarker for Donor Lung Assessment in Clinical Ex Vivo Lung Perfusion Sciencedirect:S1053249815000054
J Neurochem. Endothelin-converting enzymes degrade α-synuclein and are reduced in dementia with Lewy bodies. pubmed:28171705
Catalog No. Related products for research use of Homo sapiens (Human) Organism species Applications (RESEARCH USE ONLY!)
RPA483Hu01 Recombinant Endothelin Converting Enzyme 1 (ECE1) Positive Control; Immunogen; SDS-PAGE; WB.
PAA483Hu01 Polyclonal Antibody to Endothelin Converting Enzyme 1 (ECE1) WB; IHC; ICC; IP.
MAA483Hu22 Monoclonal Antibody to Endothelin Converting Enzyme 1 (ECE1) WB; IHC; ICC; IP.
SEA483Hu ELISA Kit for Endothelin Converting Enzyme 1 (ECE1) Enzyme-linked immunosorbent assay for Antigen Detection.
LMA483Hu Multiplex Assay Kit for Endothelin Converting Enzyme 1 (ECE1) ,etc. by FLIA (Flow Luminescence Immunoassay) FLIA Kit for Antigen Detection.