ELISA Kit for Coagulation Factor XII (F12)

FXII; HAF; Contact Factor; Hageman Factor

Specificity

This assay has high sensitivity and excellent specificity for detection of Coagulation Factor XII (F12).
No significant cross-reactivity or interference between Coagulation Factor XII (F12) and analogues was observed.

Recovery

Matrices listed below were spiked with certain level of recombinant Coagulation Factor XII (F12) and the recovery rates were calculated by comparing the measured value to the expected amount of Coagulation Factor XII (F12) in samples.

Matrix Recovery range (%) Average(%)
sodium citrate plasma(n=5) 98-105 101

Precision

Intra-assay Precision (Precision within an assay): 3 samples with low, middle and high level Coagulation Factor XII (F12) were tested 20 times on one plate, respectively.
Inter-assay Precision (Precision between assays): 3 samples with low, middle and high level Coagulation Factor XII (F12) were tested on 3 different plates, 8 replicates in each plate.
CV(%) = SD/meanX100
Intra-Assay: CV<10%
Inter-Assay: CV<12%

Linearity

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of Coagulation Factor XII (F12) and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.

Sample 1:2 1:4 1:8 1:16
sodium citrate plasma(n=5) 92-99% 78-89% 78-105% 87-101%

Stability

The stability of kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage condition.
To minimize extra influence on the performance, operation procedures and lab conditions, especially room temperature, air humidity, incubator temperature should be strictly controlled. It is also strongly suggested that the whole assay is performed by the same operator from the beginning to the end.

Reagents and materials provided

Reagents Quantity Reagents Quantity
Pre-coated, ready to use 96-well strip plate 1 Plate sealer for 96 wells 4
Standard 2 Standard Diluent 1×20mL
Detection Reagent A 1×120µL Assay Diluent A 1×12mL
Detection Reagent B 1×120µL Assay Diluent B 1×12mL
TMB Substrate 1×9mL Stop Solution 1×6mL
Wash Buffer (30 × concentrate) 1×20mL Instruction manual 1

Assay procedure summary

1. Prepare all reagents, samples and standards;
2. Add 100µL standard or sample to each well. Incubate 1 hours at 37°C;
3. Aspirate and add 100µL prepared Detection Reagent A. Incubate 1 hour at 37°C;
4. Aspirate and wash 3 times;
5. Add 100µL prepared Detection Reagent B. Incubate 30 minutes at 37°C;
6. Aspirate and wash 5 times;
7. Add 90µL Substrate Solution. Incubate 10-20 minutes at 37°C;
8. Add 50µL Stop Solution. Read at 450nm immediately.

GIVEAWAYS

INCREMENT SERVICES

Magazine Citations
Perfusion Ex vivo simulation of cardiopulmonary bypass with human blood for hemocompatibility testing PubMed: 26243277
Catalog No. Related products for research use of Mus musculus (Mouse) Organism species Applications (RESEARCH USE ONLY!)
EPA677Mu61 Eukaryotic Coagulation Factor XII (F12) Positive Control; Immunogen; SDS-PAGE; WB.
RPA677Mu02 Recombinant Coagulation Factor XII (F12) Positive Control; Immunogen; SDS-PAGE; WB.
RPA677Mu01 Recombinant Coagulation Factor XII (F12) Positive Control; Immunogen; SDS-PAGE; WB.
APA677Mu61 Active Coagulation Factor XII (F12) Cell culture; Activity Assays.
PAA677Mu01 Polyclonal Antibody to Coagulation Factor XII (F12) WB; IHC; ICC; IP.
PAA677Mu02 Polyclonal Antibody to Coagulation Factor XII (F12) WB; IHC; ICC; IP.
SEA677Mu ELISA Kit for Coagulation Factor XII (F12) Enzyme-linked immunosorbent assay for Antigen Detection.
LMA677Mu Multiplex Assay Kit for Coagulation Factor XII (F12) ,etc. by FLIA (Flow Luminescence Immunoassay) FLIA Kit for Antigen Detection.