ELISA Kit for Heat Shock Protein 60 (Hsp60)

HSPD1; GROEL; CPN60; HuCHA60; Heat Shock 60kD Protein 1, Chaperonin; Spastic Paraplegia 13,Autosomal Dominant; 60 kDa chaperonin; Mitochondrial matrix protein P1; P60 lymphocyte protein

Specificity

This assay has high sensitivity and excellent specificity for detection of Heat Shock Protein 60 (Hsp60).
No significant cross-reactivity or interference between Heat Shock Protein 60 (Hsp60) and analogues was observed.

Recovery

Matrices listed below were spiked with certain level of recombinant Heat Shock Protein 60 (Hsp60) and the recovery rates were calculated by comparing the measured value to the expected amount of Heat Shock Protein 60 (Hsp60) in samples.

Matrix Recovery range (%) Average(%)
serum(n=5) 94-105 99
EDTA plasma(n=5) 81-99 87
heparin plasma(n=5) 94-101 97

Precision

Intra-assay Precision (Precision within an assay): 3 samples with low, middle and high level Heat Shock Protein 60 (Hsp60) were tested 20 times on one plate, respectively.
Inter-assay Precision (Precision between assays): 3 samples with low, middle and high level Heat Shock Protein 60 (Hsp60) were tested on 3 different plates, 8 replicates in each plate.
CV(%) = SD/meanX100
Intra-Assay: CV<10%
Inter-Assay: CV<12%

Linearity

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of Heat Shock Protein 60 (Hsp60) and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.

Sample 1:2 1:4 1:8 1:16
serum(n=5) 78-94% 80-92% 83-99% 91-102%
EDTA plasma(n=5) 81-98% 88-95% 80-90% 99-105%
heparin plasma(n=5) 79-89% 78-97% 98-105% 97-104%

Stability

The stability of kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage condition.
To minimize extra influence on the performance, operation procedures and lab conditions, especially room temperature, air humidity, incubator temperature should be strictly controlled. It is also strongly suggested that the whole assay is performed by the same operator from the beginning to the end.

Reagents and materials provided

Reagents Quantity Reagents Quantity
Pre-coated, ready to use 96-well strip plate 1 Plate sealer for 96 wells 4
Standard 2 Standard Diluent 1×20mL
Detection Reagent A 1×120µL Assay Diluent A 1×12mL
Detection Reagent B 1×120µL Assay Diluent B 1×12mL
TMB Substrate 1×9mL Stop Solution 1×6mL
Wash Buffer (30 × concentrate) 1×20mL Instruction manual 1

Assay procedure summary

1. Prepare all reagents, samples and standards;
2. Add 100µL standard or sample to each well. Incubate 1 hours at 37°C;
3. Aspirate and add 100µL prepared Detection Reagent A. Incubate 1 hour at 37°C;
4. Aspirate and wash 3 times;
5. Add 100µL prepared Detection Reagent B. Incubate 30 minutes at 37°C;
6. Aspirate and wash 5 times;
7. Add 90µL Substrate Solution. Incubate 10-20 minutes at 37°C;
8. Add 50µL Stop Solution. Read at 450nm immediately.

GIVEAWAYS

INCREMENT SERVICES

Magazine Citations
Sci Rep Hsp60 response in experimental and human temporal lobe epilepsy PubMed: 25801186
Cellular and Molecular Gastroenterology and Hepatology Senescent hepatocytes in decompensated liver show reduced UPR and its key player, CLPP, attenuates senescence in vitro Pubmed: 30878663
Breast Cancer Res Treat Activation of mitochondrial unfolded protein response is associated with Her2-overexpression breast cancer Pubmed: 32601970
Catalog No. Related products for research use of Oryctolagus cuniculus (Rabbit) Organism species Applications (RESEARCH USE ONLY!)
SEA822Rb ELISA Kit for Heat Shock Protein 60 (Hsp60) Enzyme-linked immunosorbent assay for Antigen Detection.
LMA822Rb Multiplex Assay Kit for Heat Shock Protein 60 (Hsp60) ,etc. by FLIA (Flow Luminescence Immunoassay) FLIA Kit for Antigen Detection.