ELISA Kit for p21 Protein Activated Kinase 4 (PAK4)

Serine/threonine-protein kinase PAK 4

Specificity

This assay has high sensitivity and excellent specificity for detection of p21 Protein Activated Kinase 4 (PAK4).
No significant cross-reactivity or interference between p21 Protein Activated Kinase 4 (PAK4) and analogues was observed.

Recovery

Matrices listed below were spiked with certain level of recombinant p21 Protein Activated Kinase 4 (PAK4) and the recovery rates were calculated by comparing the measured value to the expected amount of p21 Protein Activated Kinase 4 (PAK4) in samples.

Matrix Recovery range (%) Average(%)
serum(n=5) 94-103 97
EDTA plasma(n=5) 87-103 99
heparin plasma(n=5) 84-101 93

Precision

Intra-assay Precision (Precision within an assay): 3 samples with low, middle and high level p21 Protein Activated Kinase 4 (PAK4) were tested 20 times on one plate, respectively.
Inter-assay Precision (Precision between assays): 3 samples with low, middle and high level p21 Protein Activated Kinase 4 (PAK4) were tested on 3 different plates, 8 replicates in each plate.
CV(%) = SD/meanX100
Intra-Assay: CV<10%
Inter-Assay: CV<12%

Linearity

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of p21 Protein Activated Kinase 4 (PAK4) and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.

Sample 1:2 1:4 1:8 1:16
serum(n=5) 88-97% 93-101% 81-88% 86-93%
EDTA plasma(n=5) 79-101% 96-103% 88-95% 99-105%
heparin plasma(n=5) 82-89% 80-92% 95-105% 80-101%

Stability

The stability of kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage condition.
To minimize extra influence on the performance, operation procedures and lab conditions, especially room temperature, air humidity, incubator temperature should be strictly controlled. It is also strongly suggested that the whole assay is performed by the same operator from the beginning to the end.

Reagents and materials provided

Reagents Quantity Reagents Quantity
Pre-coated, ready to use 96-well strip plate 1 Plate sealer for 96 wells 4
Standard 2 Standard Diluent 1×20mL
Detection Reagent A 1×120µL Assay Diluent A 1×12mL
Detection Reagent B 1×120µL Assay Diluent B 1×12mL
TMB Substrate 1×9mL Stop Solution 1×6mL
Wash Buffer (30 × concentrate) 1×20mL Instruction manual 1

Assay procedure summary

1. Prepare all reagents, samples and standards;
2. Add 100µL standard or sample to each well. Incubate 1 hours at 37°C;
3. Aspirate and add 100µL prepared Detection Reagent A. Incubate 1 hour at 37°C;
4. Aspirate and wash 3 times;
5. Add 100µL prepared Detection Reagent B. Incubate 30 minutes at 37°C;
6. Aspirate and wash 5 times;
7. Add 90µL Substrate Solution. Incubate 10-20 minutes at 37°C;
8. Add 50µL Stop Solution. Read at 450nm immediately.

GIVEAWAYS

INCREMENT SERVICES

Catalog No. Related products for research use of Homo sapiens (Human) Organism species Applications (RESEARCH USE ONLY!)
RPH464Hu01 Recombinant p21 Protein Activated Kinase 4 (PAK4) Positive Control; Immunogen; SDS-PAGE; WB.
PAH464Hu01 Polyclonal Antibody to p21 Protein Activated Kinase 4 (PAK4) WB
MAH464Hu21 Monoclonal Antibody to p21 Protein Activated Kinase 4 (PAK4) WB; IHC; ICC; IP.
SEH464Hu ELISA Kit for p21 Protein Activated Kinase 4 (PAK4) Enzyme-linked immunosorbent assay for Antigen Detection.
SCH464Hu CLIA Kit for p21 Protein Activated Kinase 4 (PAK4) Chemiluminescent immunoassay for Antigen Detection.
LMH464Hu Multiplex Assay Kit for p21 Protein Activated Kinase 4 (PAK4) ,etc. by FLIA (Flow Luminescence Immunoassay) FLIA Kit for Antigen Detection.